4 Operation
Model 3000MA
4-16
Teledyne Analytical Instruments
To switch from autoranging to fixed range analysis, enter the range
function by pressing the
Range
button on the front panel.
Use the < > arrow keys to move the blinking over
AUTO
.
Use the
∆∇
arrow keys to switch from
AUTO
to
FX/LO
,
FX/MED,
or
FX/HI
to set the instrument on the desired fixed range (low, medium, or
high).
L1.00 M5.00
H10.00 ModeFX/LO
or
L1.00 M5.00
H10.00 ModeFX/MED
or
L1.00 M5.00
H10.00 ModeFX/HI
Press
Escape
to re-enter the
Analyze
mode using the fixed range.
NOTE:When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc. However, the digital readout and
the RS-232 output continue to read the true value of the oxy-
gen concentration regardless of the analog output range.
4.7
The
Analyze Function
Normally, all of the functions automatically switch back to the
Analyze
function when they have completed their assigned operations. Pressing the
Escape
button in many cases also switches the analyzer back to the
Ana-
lyze
function. Alternatively, you can press the
Analyze
button at any time
to return to analyzing your sample.
4.8
Signal Output
The standard Model 3000MA Percent Oxygen Analyzer is equipped
with two 0–1 V dc analog output terminals accessible on the back panel (one
concentration and one range ID). The –MA option also has two isolated 4–
20 mA dc current outputs (one concentration and one range ID).
See Rear Panel in Chapter 3, Installation, for illustration.