7.1 Manual Zero/Span Check or Cal With Zero/Span Gas in
the Sample Port
The calibration of the instrument can be checked or adjusted using gas from the sample port.
This method is often used when the calibration gas is supplied from an external calibrator and
valve system or when the Analyzer does not include the IZS or Z/S valves option.
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero
value.
details the zero calibrate procedure with zero gas coming in through the sample
port.
Table 7-2: Manual Zero Calibration Procedure - Zero Gas Thru Sample Port
Step Number
Action
Comment
1.
Press CAL
The M200A enters the calibrate mode from sample mode. The
zero gas must come in through the sample port.
2.
Wait 10 min
Wait for reading to stabilize at zero value.
3.
Press ZERO
If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
4.
Press ENTR
Pressing ENTR actually changes the calculation equations.
5.
Press EXIT
M200A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.
Next, enter the expected NO
x
and NO span gas concentrations:
7-4
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...