
4 Operation
Model 2000A-EU
4-22
Teledyne Analytical Instruments
right corner, the abbreviation
Anlz
indicates that the analyzer is in the
Analyze
mode. If there is an * before the
Anlz
, it indicates that the range is linearized.
19.3 % O2 Air
R: ØØ:Ø 17 *Anlz
If the concentration detected is overrange, the first line of the display blinks
continuously.
4.8
Programming
CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the facto-
ry to the users application specification. These func-
tions should only be reprogrammed by trained,
qualified personnel.
To program, you must:
1. Enter the password, if you are using the analyzer’s password
protection capability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 3
Installation for details). Send the rp command to the analyzer.
OR
For software 1.1.4 or later, turn the instrument off and back on.
While on the introduction screen hold the Analyze key for at least
fifteen seconds. Press the Enter key twice to return to the
Analyze mode.
3. Press the
System
button to start the
System
function.
CONTRAST AUTOCAL
PWD LOGOUT MORE
Use the < > arrow keys to blink
MORE
, then press
Enter
.
MODEL APPLICATION
SELF_TEST ALGORITHM
Now you will be able to select the APPLICATION and ALGORITHM
set-up functions.
Contrast Function is
DISABLED
(Refer to Section 1.6)