![Teledyne SP Devices ADQ7DC Manual Download Page 40](http://html1.mh-extra.com/html/teledyne-sp-devices/adq7dc/adq7dc_manual_1081091040.webp)
19-2233 PC2
2019-02-01
40(50)
ADQ7DC Manual
16-1796 PC2 2019-02-01
40(50)
7.4 Data transfer modes
The data transfer can be set up in two main ways:
•
Event-driven streaming, see
•
User-scheduled multi-record, see
7.4.1
Streaming data transfer mode
The streaming mode is preferred for high-performance event-driven data transfer. This is the main
mode for both continuous streaming (
) and triggered streaming (
) acquisi-
tion. The flow is driven by the real-time acquisition in the digitizer. The software is set up to be prepared
to receive and process data when it arrives. Since the PC is not a real-time system, the data buffer
(
) on the ADQ is necessary. Each processing step in the software has to complete on time.
Otherwise there is a risk of overflow in the FIFO on the digitizer.