Timing Commands
The following commands set the timing of various profiling functions.
Available Timing Commands
This section lists the available Timing commands.
>T?
Available Commands:
TE 00:00:00.00 ---------- Time Between Ensembles
TP 00:05.00 ------------- Time Between Pings
TS 16/10/13,11:19:30.02 - Set System Date and Time (yy/mm/dd,hh:mm:ss)
T? ---------------------- Display T-Command Menu
>#T?
*** CAUTION: These commands are reserved for RDI use
and may not be currently supported!
Available Commands:
#TM 10 ------------------ Set minimum time between pings (ms) [0..100]
#T? --------------------- Display #T-Command Menu
>
Standard Timing Commands
TE – Time Per Ensemble
Purpose
Sets the minimum interval between data collection cycles (data ensembles).
Format
TEhh:mm:ss.ff
Range
hh
= 00 to 23 hours
mm
= 00 to 59 minutes
ss
= 00 to 59 seconds
ff
= 00 to 99 hundredths of seconds
Default
TE00:00:00.00
Recommended Setting. The default setting for this command is recommended for most applications.
Description During the ensemble interval set by TE, the Sentinel V Real-Time collects one automatic
ensemble. If TE = 00:00:00.00, the Sentinel V Real-Time starts collecting the next ensem-
ble immediately after processing the previous ensemble.
Example
TE01:15:30.00 tells the Sentinel V Real-Time to collect data ensembles every 1 hour, 15
minutes, 30 seconds.
The time tag for each ensemble is the time of the first ping of that ensemble.