WaveAce Lab 3: Making Measurements
page
|
6 of 8
Step 9:
In the CURSOR menu change Mode to Auto. For the moment nothing is displayed. This is
because in Auto mode the cursors operate in combination with a measurement parameters.
Change the timebase to 500us/div and press the MEASURE button.
To change any measurement press the corresponding button (in this case the top button).
Notes:
Parameters are pre-programmed measurements that eliminate the need to set up cursors or
standardized measurements, like rise-time, fall-time, peak-peak amplitude etc. Use them to
automatically calculate many attributes of your channel waveforms. You can display up to five selected
parameter measurements on the side menu.
Either CH1 or CH2 can be used as the source for each of the Voltage and Time measurements. For
Delay measurements both channels are required and the source setting is ignored.
Remember to turn on the source channel(s).