Teledyne Lecroy PP021 Series User Manual Download Page 10

PP021 Passive Probe 

EAR99 technology subject to restrictions on copyrights page. 

 

Over-compensated, correct, and under-compensated waveforms 

Probe Accessories 

Probe Tips 

Both the rigid probe tip and the spring tip can be screwed on to the probe 

body by hand.  

 

Be sure not to overtighten the tips to avoid damaging the threads.

 

Ground Accessories 

The ground lead and the short collar are attached to the probe at the time 

of shipping. To replace the ground lead with the swivel lead adapter, 

unscrew the short collar from the probe and remove the ground lead. 

Replace the short collar to hold the swivel lead adapter in place.  

Alternatively, the long collar can be pushed onto the probe body and used in 

conjunction with the ground spring, BNC adapter or PCB adapter. 

Summary of Contents for PP021 Series

Page 1: ...PP021 1 PP021 2 User Manual PP021 Passive Probe...

Page 2: ...ne LeCroy documentation materials is strictly prohibited Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U...

Page 3: ...oduction Teledyne LeCroy PP021 is a passive high impedance voltage oscilloscope probe designed and calibrated for use with general purpose oscilloscopes having an input impedance of 1 M It can be comp...

Page 4: ...or category ratings of the probe See voltage derating curve on p 3 PollutionDegree PollutionDegree2 OperatingAltitude Upto3000meters Max OperatingTemp 0to50 C Humidity 5 to 95 RH 10 to 30 C 5 to 75 RH...

Page 5: ...User Manual EAR99 technology subject to restrictions on copyrights page 3 PP021 Input Impedance Profile Voltage vs Frequency Derating Curve...

Page 6: ...he European Communities IEC EN 61010 031 2015 Safety requirements for electrical equipment for measurement control and laboratory use Part 031 Safety requirements for handheld probe assemblies for ele...

Page 7: ...t Disconnect the probe input and reference lead from the test circuit before disconnecting from the oscilloscope Do not connect disconnect probes while connected to a voltage source Do not overload Do...

Page 8: ...on Do not allow any portion of the probe to be submerged at any time Dry the probe and accessory thoroughly before attempting to make any voltage measurement Do not subject the probe to solvents or so...

Page 9: ...probe is connected to or transferred between oscilloscopes Low Frequency 1 Apply a 1 kHz square wave to the probe or connect to the oscilloscope s calibrator output 2 Adjust the single LF trimmer loca...

Page 10: ...ghten the tips to avoid damaging the threads Ground Accessories The ground lead and the short collar are attached to the probe at the time of shipping To replace the ground lead with the swivel lead a...

Page 11: ...Trimmer Tool 1 6 PK5 2 5MM 105 Long Collar 1 7 PK5 2 5MM 106 IC Tip Insulators Black Brown Blue Green Gray 5 8 PK5 2 5MM 107 Tip Insulator 2 5mm 1 9 PK5 2 5MM 108 Probe Tip Cover 1 10 PK5 2 5MM 109 BN...

Page 12: ...nty Teledyne LeCroy at its option will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center However this will be...

Reviews: