Teledyne API – Model T108 Addendum to T100 Manual
Instrument Test & Calibration Record
07268C DCN8258
23
Table 3-3. Test and Calibrations Values w/ CO2 where applicable
Span and Cal
Values
Acceptable Value
Parameter
Observed Value
Units
Nominal Range
TS Span Conc.
PPB
20 - 20,000
TS Slope
1.0 +/- .3
TS Offset
MV
< 100
Noise at Zero (rms)
PPB
< 0.2
Noise at Span (rms)
PPB
< 0.5
PMT at Zero (SO
2
/CO
2
)
MV
PMT at Span (SO
2
/CO
2
)
MV
Measured Flows
Parameter
Observed Value
Units
Nominal Range
Sample Flow w/ CO
2
cc/min
500 - 700
Sample Flow w/Air
cc/min
400 - 600
Sample Press w/CO
2
“ HG
24 - 27
IZS Purge Flow
cc/min
50 +/- 10
H
2
S Conversion
Efficiency
Expected = ______ PPB
Actual = ______ PPB
Efficiency = ________ %
(100
±
2%)
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Internal Zero/Span - IZS
Permeation Tube (Output Specification)
4-20 MA Current Loop Output
External Pump
PROM Rev #: _______________________
T108TS S/N: ____________
501TS S/N: ____________
Date: _______________________ Technician: _______________________