![Telect LCX Series User Manual Download Page 42](http://html.mh-extra.com/html/telect/lcx-series/lcx-series_user-manual_1080057042.webp)
Telect, Inc. • USA +1.509.926.6000 • 52.33.3836.37.52
www.telect.com • © 2010 Telect, Inc., All Rights Reserved, 126254-4 A0.1
Page 4-37
An IFC clamp consists of clamp halves that fit around the IFC.
Various sizes of IFC are accommodated using various layers
(or leaves) of bushing between the cable and clamp.
The bushing kit contains a cylinder of bushing leaves, as
shown on the right. Measure the diameter of the IFC and then
remove and discard that much from the interior of the cylinder.
The remaining bushing halves are used between the clamp
halves and the IFC, as shown in Figure 43.
6. For unprotected ribbon fiber, place ~60 in. (~1½ m) of flex tubing over exposed ribbon. The
end of the flex tubing should be pushed back to the IFC clamp.
Solid Core (Discard)
Bushing Leaf
Figure 42 - IFC Clamp
Cab le clam ps can be p iggy-backe d to allow m ore tha n
one IFC at each panel entra nce . Pig gy-backing also
provides cable clearance to lower LC X S plice Pane ls.
(Th e clam ps can also be secured to LC X P atch Pane ls.)
u
H ook an d loo p strap inserted
through tie-dow n to hold
sub units to floor of pa ne l
C able C lam p
A tt a c h c la m p p la te to
lo w e r tw o ho le s in
s id ew a ll f o r bo tt o m -up cab le f eed s .
Clamp H alf Bushing
C lamp Half
S ubunit
C la mp P late
IF
C
:1
44
S
tr
ands
in
1
2
S
ubuni
ts
Figure 43 - Installing IFC for Splicing