AFA4000/2
Airflow Monitor / p.98
7.4
Auto Sash Controller
–
stand-alone
This section refers to when the Auto Sash Controller is used without the AF4000/2/AS
Airflow Monitor.
Before starting up the Auto Sash Controller, ensure that all Auto Sash Control System
components have been correctly installed and connected.
Note: Check that the fume hood sash and associated motor drives, pulleys, operating
cables, racks and chains move freely over their entire operating ranges.
Note: Check that all cables are secured away from moving components and are not
overstressed or damaged.
Note: Check the security of all electrical connections.
7.4.1
Starting up the Auto Sash Control Unit
Press the
Power Up/Menu Entry
button to start the Auto Sash Controller (Figure 63 and
Figure 63: UK / EU keypad
Figure 64: US keypad
1.
Switch on the Auto Sash Control Unit.
When powering up, the Auto Sash Control Unit performs a self-test of its functions,
LEDs and the audible alarm. This self-test takes approximately 3 seconds.
During the self-test period, all inputs and outputs are de-activated.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...