AFA4000/2
Airflow Monitor / p.43
5.
Installation: Airflow Monitor
This section outlines the installation of the various components of the airflow monitor
system. The size and format of individual fume cupboards varies considerably, so specific
instructions are not possible, however the principles outlined are valid for all cases.
5.1
Location
The AFA4000/2 monitor can be mounted on either side of the fume cupboard. However,
when deciding on the location for the airflow monitor and sensor, consider the cable
lengths required:
•
The 5 metre low-voltage cable must reach from the terminals on the back of the
AFA4000/2 airflow monitor to the connections on top of the fume cupboard.
•
The standard 2 metre ‘telephone style’ sensor cable must reach fr
om the back of the
AFA4000/2 airflow monitor to the airflow sensor. When possible ensure that the
airflow sensor is mounted on the same side of the fume cupboard as the AFA4000/2.
Note: Longer sensor cables are available on request.
Summary of Contents for AFA4000/2
Page 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Page 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Page 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Page 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Page 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Page 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Page 127: ...AFA4000 2 Airflow Monitor p 127 ...
Page 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Page 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...