Calibration — Type Z
connectors. Make certain the VAR. ATTEN. control is set
fully clockwise. Set the G A IN ADJUST control (R8639) to
obtain exactly 4 cm of vertical deflection. Place the Mode
switch to — B O N LY. Check to see that there are exactly 4
cm of vertical deflection.
8. O perational Checks
Perform the following Z Unit operational checks:
(a) Disconnect the calibrator signal from the input con
nectors.
Check to see that the baseline does not shift
more than 2 mm after switching the Mode switch from
A O N LY to — B O N LY. Reconnect the calibrator signal
to the input connectors.
(b) With the Mode switch at — B O N LY, check the opera
tion of the B AC-DC switch by setting it to the AC posi
tion. Observe the signal to see that it shifts to its average
voltage level. The shifting of the signal indicates that the
AC-DC switch and the blocking capacitor function properly.
Set the Mode switch to A O N LY and check the A AC-DC
switch for the same waveform shift indication. Place both
AC-DC switches back to the DC position.
(c) Check operation of the VAR. ATTEN. control. When
the control is rotated fully counterclockwise, the vertical de
flection of the signal should not exceed 1.6 cm. Rotate the
control back to the fully clockwise position.
9. Turret Attenuation Ratios
With the Mode switch at A O N LY, check for 2 cm of
vertical deflection at each position of the A VOLTS/CM
switch by following the information in Table 5-2.
TABLE 5-2
VO LTS/CM Switch
Setting
Calibrator Output
Volts
.05
.1
.1
.2
.25
.5
.5
1
1
2
2.5
5
5
10
10
20
25
50
After checking the A VOLTS/CM turret attenuator, set
the Mode switch to — B O N LY and check the B VOLTS/CM
attenuator in the same manner.
10.
Differential Balance
Set the Mode switch to the A-B position and both
VOLTS/CM switches to .05. Set the calibrator switch for
a 100-volt output signal. Set the oscilloscope triggering
controls for a stable display. Ignoring the spikes on the
waveform, adjust the DIFF. BAL. control for minimum square-
wave amplitude.
The maximum allowable amplitude is
0.5 mm. When the control is correctly adjusted and the Z
Unit balances properly, the waveform appears as a straight
line with positive and negative spikes.
An ac mismatch between two tubes or components can
cause the straight line segments of the waveform to tilt
diagonally to form a zigzag pattern. The amplitude of the
tilt at its highest point should not exceed 1 mm. To check
the amplitude, free run the sweep at 20 millisec/cm and
observe the width of the trace. It should not exceed 1.5 mm
(amplitude of tilt and square wave together). After com
pleting this check, repeat step 5. Then go on to step 11.
11. Characteristic Unbalance
Disconnect the signal from the Z Unit inputs, and set
the Mode switch to the TEST position. Set the oscilloscope
Time/cm switch to .5 millisec/cm. W hile observing the
trace, move the COMPARISON VO LTAG E Polarity switch
to the + , 0, and — positions several times. The overall
baseline shift should not exceed 1 cm. Leave the CO M
PARISON V O LTAG E Polarity switch in the 0 position at
the conclusion of this step.
12. Input-Capacitance Standardization
Loosen the set screws in both VOLTS/CM knobs with a
0.05" alien wrench and remove the knobs. This allows ac
cess to the attenuator adjustments through slots in the front
panel.
Each position of the VOLTS/CM control has a
SHUNT and a SERIES adjustment. The only exception is the
.05 position which has only an input-capacitance trimmer
accessible through the SERIES side of the slot.
Connect the 52-ohm cable to the Type 105 Square-Wave
Generator, or equivalent. Connect the other end of the
cable to the 5-to-l '1 ” pad. Connect the other end of the
“ L” pad through the 24-^/xf capacitance standardizer to
the A input connector on the Z Unit. Set the Mode switch
to A O N LY and both VOLTS/CM switches to .05.
To make this adjustment, set the square-wave generator
to produce about 3.5 cm of deflection at a frequency of
1 kc. Set the oscilloscope front-panel controls for “ double
triggering” as follows:
Triggering Mode
Trigger Slope
Trigger Level
Stability
Time/cm
AC Fast
+ Internal
0
Fully clockwise
.5 Millisec
Rotate the Variable Time/cm control slowly counterclock
wise until you obtain a stable display of overlapping square
waves as shown in Fig. 5-5.
Adjust the input-capacitance trimmer in the A VOLTS/CM
attenuator for best flat top on the displayed waveform
(see Fig. 5-5), disregarding any fast-rise leading edge spikes.
Set the Mode switch to — B O N LY. Disconnect the signal
from the A input connector and connect it to the B input.
Adjust the input-capacitance trimmer in the B VOLTS/CM
attenuator for best flat bottom, disregarding the spikes.
13. Output Voltage Divider Compensation
Increase the square-wave generator frequency to 10 kc.
Set the oscilloscope Time/cm switch to 50 jusec and re
adjust the Variable Time/cm control to obtain a stable
display of several overlapping waveforms. Adjust the out
put voltage divider compensations C7655 and C8655 for
®
5-4
Summary of Contents for TYPE Z PLIG-IN
Page 14: ...OVERALL G A IN 2 PUSH PULL I OUTPUT l I l Ci I Fig 3 1 Type Z Plug In Unit block diagram 3 0...
Page 18: ...NOTES Type Z...
Page 24: ...NOTES Type Z...
Page 32: ...NOTES...
Page 36: ...0 Parts List Type Z RIGHT SIDE 0 6 4...
Page 38: ...Parts List Type Z LEFT SIDE 6 6...
Page 44: ...NOTES...
Page 52: ......