Performance Veri
fi
cation
Test Record
Table 11: Test record
Instrument Serial Number:
Temperature:
Date of Calibration:
Certi
fi
cate Number:
RH %:
Technician:
Test
Passed
Failed
Self Test
Oscilloscope
tests
High value
(measured)
Low value
(measured)
Low limit
Test result
(High value –
low value)
High limit
5 mV/div
33.6 mV
36.4 mV
200 mV/div
1.358 V
1.442 V
Channel 1
DC Gain
Accuracy
2 V/div
13.58 V
14.42 V
5 mV/div
33.6 mV
36.4 mV
200 mV/div
1.358 V
1.442 V
Channel 2
DC Gain
Accuracy
2 V/div
13.58 V
14.42 V
5 mV/div
33.6 mV
36.4 mV
200 mV/div
1.358 V
1.442 V
Channel 3
DC Gain
Accuracy
2 V/div
13.58 V
14.42 V
5 mV/div
33.6 mV
36.4 mV
200 mV/div
1.358 V
1.442 V
Channel 4
DC Gain
Accuracy
2 V/div
13.58 V
14.42 V
20
TPS2000B Series Digital Storage Oscilloscopes Service Manual
Summary of Contents for TPS2000B Series
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscopes ZZZ Service Manual P077044701 077 0447 01...
Page 2: ......
Page 12: ...Service Safety Summary viii TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 24: ...Specifications 10 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 26: ...Operating Information 12 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 32: ...Theory of Operation 18 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 46: ...Performance Verification 32 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 86: ...Diagrams 72 TPS2000B Series Digital Storage Oscilloscopes Service Manual...