Speci
fi
cations
Mechanical Characteristics
Table 8: Mechanical characteristics
Characteristic
Description
Instrument only
2.7 kg (6.0 lbs)
1 battery installed
3.2 kg (7.0 lbs)
Weight, approximate
2 batteries installed
3.7 kg (8.0 lbs)
Height
161 mm (6.33 in)
Width
336 mm (13.24 in)
Dimensions, without
front protective cover
Depth
130 mm (5.1 in)
Height
164 mm (6.45 in)
Width
347 mm (13.65 in)
Dimensions, with front
protective cover
Depth
135 mm (5.3 in)
Cooling method
Temperature controlled fan. Intake vents on the bottom; exhaust on the right side.
Environmental Performance
Table 9: Environmental performance
Characteristic
Description
Operating
Nonoperating
Temperature
0°C to 50 °C (32 °F to 122 °F),
with 5°C/minute maximum gradient,
noncondensing, up to 3000 altitude.
-40 °C to +71 °C (-40 °F to 159.8 °F), with 5°C/minute
maximum gradient.
Operating
Nonoperating
5% to 90% relative humidity (%RH) at up
to +30 °C (86 °F )
5% to 90% relative humidity (%RH) at up to +30 °C
(86 °F)
Humidity
5% to 60% relative humidity (%RH) at up
to +50 °C (122 °F)
5% to 60% relative humidity (%RH) b30 °C
(86 °F) and +71 °C (160 °F) noncondensing.
Altitude, Operating and
Nonoperating
3,000 m (10,000 ft)
Altitude is limited by possible damage to the LCD at higher altitudes. This damage is independent of
operation.
Pollution Degree
2
Operating
Nonoperating
Random Vibration
0.31 g
RMS
from 5 Hz to 500 Hz, 10 minutes
on each axis, 3 axes (30 minutes total).
2.46 g
RMS
from 5 Hz to 500 Hz, 10 minutes on each axis,
3 axes (30 minutes total).
Mechanical Shock,
operating
4
Half-sine mechanical shocks, 50 g peak amplitude, 11 ms duration, 3 drops in each direction of each
axis (18 total)
1
Charge battery packs between 0 °C and 45 °C (32 °F to 113 °F). Charging battery packs outside this range can damage cells or cause them to leak.
2
Battery packs are rated to operate between -10 °C and +50 °C (-14 °F and +122 °F) with less than 80% relative humidity. Operating outside this range can
cause damage. Battery discharge capacity drops signi
fi
cantly at temperatures below 0 °C (32 °F) and above 45 °C (113 °F).
3
Max wet bulb.
4
With one battery pack installed.
TPS2000B Series Digital Storage Oscilloscopes Service Manual
9
Summary of Contents for TPS2000B Series
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscopes ZZZ Service Manual P077044701 077 0447 01...
Page 2: ......
Page 12: ...Service Safety Summary viii TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 24: ...Specifications 10 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 26: ...Operating Information 12 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 32: ...Theory of Operation 18 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 46: ...Performance Verification 32 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 86: ...Diagrams 72 TPS2000B Series Digital Storage Oscilloscopes Service Manual...