Speci
fi
cations
These speci
fi
cations apply to all TPS2000B Series oscilloscopes. To verify that an
oscilloscope meets speci
fi
cations, it must
fi
rst meet the following conditions:
The oscilloscope must have been operating continuously for twenty minutes
within the speci
fi
ed operating temperature range. (See Table 9.)
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
The oscilloscope must be within the factory calibration interval of one year.
The V/div speci
fi
cations apply with the probe calibration set to 1X probe
mode.
Speci
fi
cations begin in the following table. All speci
fi
cations are guaranteed
unless noted "typical." Speci
fi
cations that are marked with the
symbol are
checked in the
Performance Veri
fi
cation
section.
Signal Acquisition System
Table 1: Signal acquisition system
Characteristic
Description
Number of Input Channels
Two (TPS2012B) or four (TPS2014B, TPS2024B) identical channels, digitized simultaneously
Input Coupling
DC, AC, or GND
AC coupling connects a capacitor in series with the input circuitry. The DC input impedance
becomes very high, since capacitance is in series with all paths to ground.
Ground coupling mode provides a reference waveform derived from the values identi
fi
ed during
SPC. This reference waveform shows visually where signal reference (ground) is expected
to be.
Input Impedance, DC Coupled
1 M
Ω
±2% in parallel with 20 pF ±3 pF
Overvoltage Category
Maximum Voltage
CAT I and CAT II
300 V
RMS
CAT III
150 V
RMS
Maximum Input Voltage (between
Signal and Reference at input
BNC)
At front panel connector, 300 V
RMS
; derate at 20 dB/decade above 100 kHz to 13 V peak
AC at 3 MHz and above.
All inputs are rated 300 V CAT II, signal to
fl
oating common.
Derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above. For
nonsinusoidal waveforms, peak value must be less than 450 V. Excursion above 300 V should
be less than 100 ms duration. RMS signal level, including any DC component removed through
AC coupling, must be limited to 300 V. If these values are exceeded, damage to the oscilloscope
may result. Refer to the Overvoltage Category description. (See Table 9 on page 9.)
Maximum Float Voltage
Rated 600 V
RMS
CAT II.
Between instrument common (chassis) and BNC shell. Using rated connectors or accessories.
Number of Digitized Bits
8-bit resolution (except when set to 2 mV/div), each channel sampled simultaneously.
TPS2000B Series Digital Storage Oscilloscopes Service Manual
1
Summary of Contents for TPS2000B Series
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscopes ZZZ Service Manual P077044701 077 0447 01...
Page 2: ......
Page 12: ...Service Safety Summary viii TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 24: ...Specifications 10 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 26: ...Operating Information 12 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 32: ...Theory of Operation 18 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 46: ...Performance Verification 32 TPS2000B Series Digital Storage Oscilloscopes Service Manual...
Page 86: ...Diagrams 72 TPS2000B Series Digital Storage Oscilloscopes Service Manual...