Table 1: Characteristics

Characteristic Description

Frequency
range

Preamp: 9 kHz to 6 GHz
MDO4XX4-6 with preamp: 50 kHz
to 6 GHz
MDO4XX4-3 with preamp: 50 kHz
to 3 GHz

Preamp
gain

Amplifying state: 12 dB (nominal)
Bypass state: –1.5 dB (nominal)

Displayed
average
noise level
(DANL)
with the
preamp
attached
to the
MDO RF
input

With the preamp mode set to
“Auto” and the reference level set
to –40 dBm:
MDO4XXX-6:
50 kHz to 5 MHz: < –140 dBm/Hz
(–144 dBm/Hz typical)
5 MHz to 3 GHz: < –158 dBm/Hz
(–162 dBm/Hz typical)
3 GHz to 6 GHz: < –150 dBm/Hz
(–153 dBm/Hz typical)
MDO4XX4-3:
50 kHz to 5 Mhz: < –140 dBm/Hz
(–144 dBm/Hz typical)
5 MHz to 3 GHz: < –158 dBm/Hz
(–162 dBm/Hz typical)

Level
measurement
uncertainty
with the
preamp
attached
to the
MDO RF
input

This speci

fi

cation is for the

MDO RF channel with a preamp
attached, input levels of –2 dBm
to –62 dBm, RBW set to “Auto”,
and applies to both amplifying and
bypass states.
< ±1.5 dBm (< ±1 dBm typical), 20
°C to 30 °C temperature range.

±2 dBm (max), over full operating

range.

Maximum
operating
input
level with
preamp
attached
to the
MDO RF
input

Average continuous power:
+30 dBm (1 W)
DC maximum before damage:
±20 V DC
Maximum power before damage
(CW): +30 dBm (1 W)
Maximum power before damage
(pulse): +45 dBm (32 W)
(<10

μ

s pulse width, <1% duty

cycle, and reference level of

+10 dBm)

Characteristic Description

Reference
level
range
with the
preamp
attached
to the
MDO RF
input

Amplifying state: –30 dBm to
DANL.
Bypass state: +30 dBm to DANL.

Connector
type

SMA – female (outside threads)

Table 2: Environmental Characteristics

Characteristic Description

Temperature

Operating: 0 °C to +50 °C
Non-operating: -20 °C to +60 °C

Humidity

Operating:
High: 40 °C to 50 °C (104 °F to
122 °F), 10% to 60% RH
Low: 0 °C to 40 °C (32 °F to
104 °F), 10% to 90% RH
Non-operating:
High: 40 °C to 60 °C (104 °F to
122 °F), 10% to 60% RH
Low: 0 °C to 40 °C (32 °F to
104 °F), 5 to 90% RH

Altitude

Operating: Up to 3,000 meters
Non-operating: Up to
12,000 meters

Power Requirements.

The TPA-N-PRE

is powered directly from the RF input on
MDO4000 Series oscilloscopes.

Regulatory.

Compliance labeling: WEEE

(European Union).

Recommended Oscilloscopes.

MDO4000 Mixed

Domain Oscilloscopes
Note: For best probe support, download and
install the latest version of the oscilloscope

fi

rmware from www.tektronix.com

Performance Veri

fi

cation

Since the TPA-N-PRE preamp is an accessory
to the MDO4000, its performance veri

fi

cation

(PV) procedure is a variation on the MDO4000
PV procedure.

To run the TPA-N-PRE PV, see the most
recent MDO4000 Series Oscilloscopes
Technical Reference (077-0583-XX) at
www.tektronix.com.

Safety Summary

To avoid potential hazards, use this product only
as speci

fi

ed.

To avoid

fi

re or personal injury, do not operate

in wet/damp conditions.

Keep product surfaces clean and dry.

Warranty Information

For warranty information, go to
www.tektronix.com\service, and then use
the provided links to search for your product’s
warranty.

Contacting Tektronix

Tektronix, Inc.
14200 SW Karl Braun Drive
PO Box 500 Beaverton, OR 97077
USA

For product information, sales, service, and
technical support:

In North America, call 1–800–833–9200.

Worldwide, visit www.tektronix.com to

fi

nd

contacts in your area.

Copyright © Tektronix, Inc. All rights reserved.

www.tektronix.com

Reviews: