background image

xx

THDP0100/0200 & TMDP0200

High Voltage Differential Probes

Installation and Safety Instructions &

Product Documentation CD

ZZZ

*P071289001*

071-2890-01

Summary of Contents for THDP0100

Page 1: ...xx THDP0100 0200 TMDP0200 High Voltage Differential Probes Installation and Safety Instructions Product Documentation CD ZZZ P071289001 071 2890 01...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...not float the output of this probe The output cable must be connected to an earth referenced 0 V oscilloscope Avoid Electric Shock When using probe accessories never exceed the lowest rating of the pr...

Page 4: ...onsult the product manual for further ratings information before making connections to the product Do not apply a potential to any terminal including the common terminal that exceeds the maximum ratin...

Page 5: ...ct Surfaces Clean and Dry Terms in this Manual These terms may appear in this manual WARNING Warning statements identify conditions or practices that could result in injury or loss of life CAUTION Cau...

Page 6: ...product DANGER indicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazar...

Page 7: ...Safety requirements for electrical equipment for measurement control and laboratory use Part 031 Safety requirements for hand held probe assemblies for electrical measurement and test Canadian Certifi...

Page 8: ...nonconductive pollution occurs Occasionally a temporary conductivity that is caused by condensation must be expected This location is a typical office home environment Temporary condensation occurs o...

Page 9: ...allations performed at the source of low voltage installation Overvoltage Category III For measurements and installations performed in the building installation Overvoltage Category II For measurement...

Page 10: ...substances into the environment and to reduce the use of natural resources we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or...

Page 11: ...AC 500 Vrms Maximum common mode voltage and input voltage to earth Vrms 2 6000 V DC peak AC 2300 V CAT I 1000 V CAT III 1500V DC peakAC 1000 V CAT II 600 V CAT III 750V DC peak AC 550 V CAT I 300 V C...

Page 12: ...Operating Considerations Figure 1 Specification table footnotes referring to the probe input limits on the probes 10...

Page 13: ...32 F to 104 F Nonoperating 30 C to 70 C 22 F to 158 F Humidity Operating 5 to 85 R H 0 C to 40 C 32 F to 104 F Nonoperating 5 to 85 R H 0 C to 40 C 32 F to 104 F 5 to 45 R H above 40 C to 70 C 104 F t...

Page 14: ...fferential Probes Instruction Manual on the documentation CD Connecting the Probe to the Oscilloscope To connect the probe to the oscilloscope align the probe to the input channel and push in the prob...

Page 15: ...the probe accessory whichever is the lesser of the two Refer to the tables on the following pages for the correct accessory to use with your probe When using multiple accessories for example the exte...

Page 16: ...st probes below The test probes accept the hook tips in the THDP0100 probe accessory kit WARNING To avoid electrical shock or fire make sure the test leads are in good condition The input leads and ex...

Page 17: ...s to browse multiple test points or to connect the extender leads to the hook tips WARNING To prevent arc flash use caution when probing circuits with raised components Avoid getting the metal shell b...

Page 18: ...and then clamp the hooks onto the circuit WARNING To reduce the risk of shock when measuring voltages above 1000 V always keep your fingers behind the tactile indicator on the TASH and TALH hook tips...

Page 19: ...000 V CAT II 600 V CAT II 550 V CAT I 300 V CAT II Large hook tips TALH 2300 V CAT I 1000 V CAT II 1000 V CAT II 600 V CAT II 550 V CAT I 300 V CAT II 1 The THDP0200 and TMDP0200 probes can also be us...

Page 20: ...ircuit either directly or through any of the standard accessories described below WARNING To avoid electrical shock or fire make sure the test leads are in good condition The input leads and extender...

Page 21: ...sulator sheath at the probe tip extends and retracts into CAT III and CAT IV rated spacings Twist the probe body past the detent at each end of the twist to lock the probe into the CAT setting Maximum...

Page 22: ...n to the threaded tips of the TP175 FL handheld probes The adapters hold the pogo pin tips included in the kit Maximum ratings 150 V CAT II 0 1 A Extended test probe adapters 012 1724 xx These adapter...

Page 23: ...atings 1000 V CAT III 600 V CAT IV Pincer clips AC283 FL These clips have a pair of finer contacts that close around leads on smaller components Maximum ratings 1000 V CAT III 600 V CAT IV Alligator c...

Page 24: ...300 V CAT I 300 V CAT I Pogo pin tip adapters and tips 020 3107 xx 150 V CAT II 150 V CAT II 150 V CAT II Extended test probe adapters 012 1724 xx 300 V CAT I 300 V CAT II 300 V CAT II 300 V CAT I 30...

Page 25: ...ements and to make hands free connections on your circuit Table 2 Additional THDP0200 and TMDP0200 probe standard accessories Item Description TPH1000 probe holder The TPH1000 probe holder allows you...

Page 26: ...ard accessories cont Item Description THV Browser The THV Browser allows you to set and lock the spacing between two handheld probe tips and then browse your circuit with one hand To make a stationary...

Page 27: ...r a maximum peak voltage of 6000 V between the inputs The inputs are CAT I rated to a maximum of 2 300 VRMS between either input and earth ground The maximum Over Voltage Transient OVT rating and othe...

Page 28: ...ake the measurement with the probe See warning below WARNING The Overrange indicator does not detect an overrange condition of common mode voltages or voltage to earth potential at the probe inputs Th...

Page 29: ...ics from the measurement 5 MHz is close to the switching frequency of most switching transistors FETs in switch mode power supplies Press the button again to return to the FULL position which selects...

Page 30: ...umentation CD included with this booklet contains pdf files of the THDP0100 0200 TMDP0200 High Voltage Differential Probes Instruction Manual and other documents to help you understand how to take dif...