Tutorial
TDSDDM1 Disk Drive Measurement Application User Manual
2–9
4. Enable hysteresis by pressing Hysteresis (main)
➞
Hysteresis (side). Using
the general purpose knob, set a hysteresis value that is appropriate for your
disk drive signal, such as 1.5 divs.
The hysteresis value, in divisions, is reflected above and below the wave-
form baseline. For example, if you choose a value of 2.0 divs, the hysteresis
level ranges from –2.0 divs to +2.0 divs.
5. Enable limit testing by pressing Limit Test (main)
➞
On/Off (side). Select
ON.
6. Press Upper Limit (side), and use the general purpose knob to select an
upper limit for this test, such as 100
m
s.
7. Press Lower Limit (side), and use the general purpose knob to select a lower
limit for this test, such as 1 ns.
8. Press –menu– (main)
➞
Control (main)
➞
Start (side) to start the test.
9. Wait for the measurement to complete, then view the results as shown in
Figure 2–7.
Figure 2–7: Pulse width at 50% (PW50) measurement with cursor gating
Summary of Contents for TDSDDM1
Page 4: ......
Page 10: ...Table of Contents vi TDSDDM1 Disk Drive Measurement Application User Manual ...
Page 13: ...Getting Started ...
Page 14: ......
Page 27: ...Operating Basics ...
Page 28: ......
Page 41: ...Reference ...
Page 42: ......
Page 46: ...Menu Structure 3 4 TDSDDM1 Disk Drive Measurement Application User Manual ...
Page 86: ...Measurement Descriptions 3 44 TDSDDM1 Disk Drive Measurement Application User Manual ...
Page 91: ...Appendices ...
Page 92: ......
Page 94: ...Appendix A Specifications A 2 TDSDDM1 Disk Drive Measurement Application User Manual ...
Page 95: ...Index ...
Page 96: ......
Page 100: ...Index Index 4 TDSDDM1 Disk Drive Measurement Application User Manual ...
Page 101: ......
Page 102: ......