background image

Copyright

Tektronix, Inc.

Beaverton, OR 97077

P6243 or P6245

Connect the probe tip to the short pin and the

probe ground to the long pin as shown.

P6246, P6247, P6248, or P7330

Connect the probe + input to the signal pin and

the probe -- input to the ground pin as shown.

There is no connection to the probe ground input.

P7330

with

adapter

P6249 or P7240

Connect the probe tip to the short pin and the

probe ground to the long pin as shown.

TCP-202

Snap the current probe closed around the current

loop as shown. Be sure to align the polarity arrows

on the current probe and the fixture.

CT-6

Remove the small jumper on the A side of the

fixture. Slip the CT-6 probe head over the

outboard pin, and then replace the jumper.

After you are finished with this probe, be sure to

replace the small jumper.

Using the Fixture

Do these steps to use the fixture. For best results,

warm up the oscilloscope for 20 minutes before

performing these functions.

Low Frequency Compensation

1. Connect fixture to the oscilloscope.
2. Connect one probe to the fixture.
3. Push the

AUTOSET

button on oscilloscope.

4. Optimize the adjustment on the probe

compensation box for best square corner.

5. Remove the connections.

Optimize Gain and Offset Accuracy

1. Connect the A side of the fixture to the

oscilloscope.

2. Remove the small jumper from the fixture.
3. Connect one probe to the A side of the fixture.
4. Touch the

VERT

button or use the

Vertical

menu to display the vertical control window.

5. Touch the Probe

Cal

button to display the

vertical probe cal control window.

6. Select the oscilloscope channel to which the

probe is attached.

7. Touch the

Calibrate Probe

button.

8. When the routine is finished, remove the

connections and replace the small jumper.

Compensate Probe Timing (Deskew)

1. Connect fixture to the oscilloscope.
2. Connect up to four probes to the fixture.
3. Display all channels that you want to deskew.
4. Push the

AUTOSET

button on oscilloscope.

5. Adjust vertical

SCALE

, and

POSITION

(with active probes adjusting

offset

may be

required) for each channel so that the signals

overlap and are centered on screen.

6. Adjust horizontal

POSITION

so that a

triggered rising edge is at center screen.

7. Adjust horizontal

SCALE

so that the

differences in the channel delays are clearly

visible.

8. Adjust horizontal

POSITION

again so that

the first rising edge is exactly at center screen.

The fastest probe is connected to this channel.

(The fastest probe is probably the one with the

shortest cable or the highest bandwidth.)

9. Touch the

VERT

button or use the

Vertical

menu to display the vertical control window.

10. Touch the Probe

Deskew

button to display the

channel-deskew control window.

11. Select one of the slower channels.

NOTE.

If possible, do the next step at a signal

amplitude within the same attenuator range

(vertical scale) as your planned signal

measurements. Any change to the vertical

scale after deskew is complete may introduce

a new attenuation level (you can generally

hear attenuator settings change) and

therefore a slightly different signal path. This

different path may cause a 200 ps variation

in timing accuracy between channels.

12. Adjust the deskew time for that channel so

that its signal aligns with that of the fastest

channel.

13. Repeat steps 11 and 12 for each additional

channel that you want to deskew.

14. Remove the connections.

Reviews: