background image

067-0405-01 and above

TDS7000 Series

Probe Calibration and

Deskew Fixture

Instructions

071-0730-03

Safety Summary

To avoid potential hazards, use this product only

as specified.

To avoid fire or personal injury

Do not operate in wet/damp conditions.
Keep product surfaces clean and dry.

Safety terms in this manual

WARNING.

These statements identify

conditions or practices that could result in

personal injury or loss of life.

Overview

This fixture converts the TDS7000 Series PROBE

COMPENSATION output into a set of test points

that give you a convenient way to:

H

Compensate a passive probe (low-frequency).

H

Optimize the oscilloscope gain and offset

accuracy at the probe tip.

H

Compensate for timing differences between

two, three, or four voltage probes (deskew).

H

Compensate for timing differences between
one, two, or three voltage probes and a current

probe (deskew).

Connecting to the Oscilloscope

The fixture has two BNC inputs, A and B. Use the

BNC cable (included) to connect the oscilloscope

PROBE COMPENSATION output to either the A

or B input. The input you use depends on the type

of probes you have and the operation you want to

perform.

Probe

2

Operation

Use

input

Passive

Low frequency compensation

A

Passive

voltage probes,

including

Gain and offset calibration

1

A

including

P6139A

Compensate timing with

respect to other voltage probes

A

Compensate timing with

respect to TCP202 current

probe

B

Compensate timing with

respect to CT-6 current probe

A

Active voltage

probes

Gain and offset calibration

A

probes,

including

P6243

P6245

Compensate timing with

respect to other voltage probes

A

P6245

P6246

P6247

P6248

P6249

Compensate timing with

respect to TCP202 current

probe

B

P6249

P7240

P7330

Compensate timing with

respect to CT-6 current probe

A

Current probe

TCP202

Compensate timing with

respect to voltage probes

B

Current probe

CT-6

Compensate timing with

respect to voltage probes

A

1

Gain and offset calibration for these probes is

generally not needed.

2

Not all probes are useable with all instruments.

NOTE.

The A side of the fixture has a small

jumper over two pins. Remove the jumper for

gain and offset calibration. Install the

jumper for all other functions.

Connecting Probes

WARNING.

To avoid personal injury, use care

while connecting probe tips to the square

pins on the fixture. The ends of the square

pins are sharp.

The following diagrams show how to attach each

type of probe tip to the fixture. Refer to the

ground symbols screened on the fixture to

establish the correct polarity.

P6139A

Connect the probe tip and the ground lead to the

two terminals as shown.

B side

connection

A side

connection

*P071073003*