TDS3000C Series Oscilloscopes
Getting Started
To optimize your measurements, it is important
to start at the beginning – by compensating
your probes and returning your oscilloscope to
a known state.
To compensate your probe, follow these
simple steps:
1. Connect the probe to channel 1.
2. Attach the probe tip and reference lead to the
Probe Comp
connectors. If using the probe hook-tip,
ensure a proper connection by firmly twisting the tip
onto the probe.
3. Push
Autoset
.
4. Check the shape of the
displayed waveform to
determine if your probe
is compensated correctly.
5. If necessary, adjust your probe. Repeat as needed.
To return your oscilloscope to a known state,
follow these simple steps:
1. Press the front panel
Save/Recall
menu button.
2. Press the
Recall Factory Setup
bottom screen
menu button.
3. Press the
OK Confirm Factory INIT
side screen
menu button.
4. Press the front panel
Autoset
button.
Autoset
button
Probe
Comp
Channel 1
Adjusting Probe P6139A
Overcompensated
Undercompensated
Compensated Correctly