![Tektronix TDS2000C series Manual Download Page 15](http://html1.mh-extra.com/html/tektronix/tds2000c-series/tds2000c-series_manual_1079157015.webp)
TDS2000C Series Oscilloscope
www.tektronix.com/tds2000
15
The TDS2000C Series offers a standard FFT feature, making it an
ideal tool for these identifying sources of noise in a circuit. To create
FFT display:
1. Press the front panel
MATH
button.
2. Press the
Operation
menu button until
FFT
is selected.
3. Press the
Window
menu button until the
Hanning
window is
selected, which provides the highest frequency resolution.
4. If desired, use the vertical and horizontal
POSITION
and
SCALE
controls to magnify and position the FFT waveform.
5. Cursors can be used to accurately measure the FFT waveform.
Press
CURSOR
.
6. Press the
Source
menu button until
MATH
is selected.
7. Press the
Type
menu button until
Frequency
is selected.
8. Press the
Cursor 1
menu button. Using the multipurpose knob,
position cursor 1 at the left side of the display.
9. Press the
Cursor 2
menu button. Position cursor 2 on the highest
point in the display.
10. The readout indicates one of the sources of noise. In this case,
the 20 MHz signal is the system clock, which is coupling into the
signal.
Look for Unintentional Circuit Noise (continued)