Performance Tests
1–66
TDS Family Performance Verification and Specifications
Figure 1–25: Measurement of Probe Compensator Amplitude
g. Check against limits:
H
Subtract the value just obtained (base level) from that obtained
previously (top level).
H
CHECK that the difference obtained is within 495 mV to 505 mV,
inclusive.
H
Enter voltage difference on test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.