Technical Reference

TDS 520B, TDS 540B, TDS 620B, TDS 644B, 
TDS 680B, TDS 684B, TDS 724A, TDS 744A, 
& TDS 784A
Digitizing Oscilloscopes
Performance Verification and Specifications

070-9384-01

Summary of Contents for TDS 520B

Page 1: ...Technical Reference TDS 520B TDS 540B TDS 620B TDS 644B TDS 680B TDS 684B TDS 724A TDS 744A TDS 784A Digitizing Oscilloscopes Performance Verification and Specifications 070 9384 01...

Page 2: ...tents issued and pending Information in this publication supercedes that in all previously published material Specifications and price change privileges reserved Printed in the U S A Tektronix Inc P O...

Page 3: ...any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obligated to furnish ser...

Page 4: ......

Page 5: ...Record 1 23 Signal Acquisition System Checks 1 27 Time Base System Checks 1 43 Trigger System Checks 1 45 Output Signal Checks 1 58 Option 05 Video Trigger Checks 1 67 Sine Wave Generator Leveling Pr...

Page 6: ...Table of Contents ii TDS Family Performance Verification and Specifications...

Page 7: ...terminal that is outside the range specified for that terminal This product is grounded through the grounding conductor of the power cord To avoid electric shock the grounding conductor must be conne...

Page 8: ...pear in this manual WARNING Warning statements identify conditions or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could resul...

Page 9: ...oltage Protective Ground Earth Terminal ATTENTION Refer to Manual Double Insulated Certifications and Compliances CSA Certification includes the products and power cords appropriate for use in the Nor...

Page 10: ...General Safety Summary vi TDS Family Performance Verification and Specifications...

Page 11: ...rvice of the digitizing oscilloscope H The TDS 520B 540B 620B 644B 680B 684B 724A 744A 784A User Manual Tektronix part number 070 9383 XX H The TDS Family Programmer Manual Tektronix part number 070 9...

Page 12: ...Preface viii TDS Family Performance Verification and Specifications...

Page 13: ...Performance Verification Procedures...

Page 14: ......

Page 15: ...Functional Tests that begin on page 1 7 Advantages These procedures require minimal additional time to perform require no additional equipment other than a standard accessory probe and more completel...

Page 16: ...ep below Initialize the oscilloscope by doing Press save recall SETUP Now press the main menu button Initialize the oscilloscope Press save recall SETUP Now press the main menu button Recall Factory S...

Page 17: ...of waveform record relative to the screen and display General purpose knob readout Waveform reference symbols show ground levels and waveform sources Graticule and waveforms Side menu area Readouts fo...

Page 18: ...Performance Verification Procedures 1 4 TDS Family Performance Verification and Specifications...

Page 19: ...is procedure 1 Verify that internal diagnostics pass Do the following substeps to verify passing of internal diagnostics a Display the System diagnostics menu H Press SHIFT then press UTILITY H Repeat...

Page 20: ...ompensation Press the main menu button Signal Path then press the side menu button OK Compensate Signal Paths g Wait Signal path compensation may take five minutes on the TDS 500B 700A and fifteen min...

Page 21: ...d probe STOP These procedures verify functions that is they verify that the oscilloscope features operate They do not verify that they operate within limits Therefore when the instructions in the func...

Page 22: ...o PROBE COMPENSATION SIGNAL on the front panel connect the probe ground typically black to PROBE COMPENSA TION GND If using a P6243 or P6245 probe you may want to attach a Y lead connector and two SMD...

Page 23: ...Set the horizontal SCALE to 200 s Press CLEAR MENU to remove any menu that may be on the screen d Verify that the channel is operational Confirm that the following statements are true H The vertical...

Page 24: ...til all four input channels are verified 3 Remove the test hookup Disconnect the probe from the channel input and the probe compensation terminals Equipment Required One probe such as the P6243 P6245...

Page 25: ...robe compensation terminals Equipment Required One probe such as the P6243 P6245 or P6139A Prerequisites None 1 Install the test hookup and preset the oscilloscope controls a Hook up the signal source...

Page 26: ...riggerable then press the side menu button Delayed Only H Set the horizontal SCALE for the D delayed time base to 200 s b Select the delayed trigger level menu H Press SHIFT then press DELAYED TRIG H...

Page 27: ...the probe on CH 1 Connect the probe tip to PROBE COMPENSATION SIGNAL on the front panel connect the probe ground to PROBE COMPENSATION GND See Figure 1 3 on page 1 8 b Insert the test disk Insert the...

Page 28: ...ll Factory Setup and the side menu button OK Confirm Factory Init to restore the 500 s time base and the five cycle waveform H Press the main menu button Recall Saved Setup then press the side menu bu...

Page 29: ...valid confirmation of perform ance and functionality when the following requirements are met H The cabinet must be installed on the Digitizing Oscilloscope H You must have performed and passed the pr...

Page 30: ...Connector Dual Banana two required Female BNC to dual banana Tektronix part number 103 0090 00 Various Accuracy Tests 7 Connector BNC T Male BNC to dual female BNC Tektronix part number 103 0030 00 Ch...

Page 31: ...plitter Power Frequency range DC to 1 GHz Tracking 2 0 Rohde Schwarz RVZ3 Checking Analog Bandwidth 22 Generator Function Frequency range 5 MHz to 10 MHz Square wave transition time 25 ns Amplitude ra...

Page 32: ...Performance Tests 1 18 TDS Family Performance Verification and Specifications...

Page 33: ...5 mV 69 1 mV 691 mV __________ __________ __________ __________ __________ __________ 1 45 mV 69 1 mV 691 mV DC Voltage Measurement Accuracy Averaged CH1 5 mV Vert scale setting 5 Div position setting...

Page 34: ...n setting 10 V offset 11 786 V __________ __________ 11 420 V CH3 1 V Vert scale setting or AX1 5 Div position setting 10 V offset 17 26 V __________ __________ 18 76 V CH3 1 V Vert scale setting or A...

Page 35: ...10 137 V Main Trigger DC Coupled Negative Slope 9 863 V __________ __________ 10 137 V Delayed Trigger DC Coupled Positive Slope 9 863 V __________ __________ 10 137 V Delayed Trigger DC Coupled Nega...

Page 36: ...bration Technician TDS 600B Performance Test Maximum Outgoing Incoming Minimum CH 3 or AX1 SIGNAL OUTPUT 1 M Pk Pk 80 mV __________ __________ Pk Pk 120 mV CH 3 or AX1 SIGNAL OUTPUT 50 Pk Pk 40 mV ___...

Page 37: ...mV 25 1 mV 251 mV __________ __________ __________ __________ __________ __________ 1 6 mV 25 1 mV 251 mV DC Voltage Measurement Accuracy Averaged CH1 5 mV Vert scale setting 5 Div position setting 1...

Page 38: ...n setting 10 V offset 11 6615 V __________ __________ 11 5385 V CH3 1 V Vert scale setting or AX1 5 Div position setting 10 V offset 17 7785 V __________ __________ 18 2215 V CH3 1 V Vert scale settin...

Page 39: ...Trigger DC Coupled Negative Slope 9 9393 V __________ __________ 10 1147 V Delayed Trigger DC Coupled Positive Slope 9 9393 V __________ __________ 10 1147 V Delayed Trigger DC Coupled Negative Slope...

Page 40: ...ration Technician TDS 500B 700A Performance Test Maximum Outgoing Incoming Minimum CH 3 or AX 1 SIGNAL OUTPUT 1 M Pk Pk 88 mV __________ __________ Pk Pk 132 mV CH 3 or AX 1 SIGNAL OUTPUT 50 Pk Pk 44...

Page 41: ...default settings H Press SHIFT then press ACQUIRE MENU H On the TDS 600B press the main menu button Mode then press the side menu button Average 16 H On the TDS 500B and 700A press the main menu butt...

Page 42: ...51 mV 1 Vertical position is set to 0 divisions and vertical offset to 0 V when the oscilloscope is initialized in step 1 c Display the test signal The waveform position and offset were initialized fo...

Page 43: ...hookup was required WARNING The generator is capable of outputting dangerous voltages Be sure to set the DC calibration generator to 0 volts before connecting disconnecting and or moving the test hook...

Page 44: ...he other side of the BNC T connector Now connect the BNC T connector to CH 1 See Figure 1 5 DC Calibrator 50 Coaxial Cables Dual Banana to BNC Adapters BNC T Connector Digitizing Oscilloscope Output S...

Page 45: ...e vertical SCALE to one of the settings listed in Table 1 3 that is not yet checked Start with the first setting listed Table 1 3 DC Accuracy Scale Setting Position Setting Divs Offset Setting Generat...

Page 46: ...the test signal Press CLEAR MENU Read the measurement results at the Mean measurement readout See Figure 1 6 Turn on the measurement called mean and read the results here Figure 1 6 Measurement of DC...

Page 47: ...o 50 precision cables Item 5 Two 10X attenuators Item 1 Optional One high frequency leveled sine wave generator and its leveling head Item 26 replaces items 19 20 21 23 24 and 5 Prerequisites See page...

Page 48: ...generator with a leveling head item 26 see Figure 1 8 and if using this optional setup with the example Tektronix SG 504 set the generator output to 6 MHz Digitizing Oscilloscope Sine Wave Generator...

Page 49: ...cal Scale Reference Amplitude Horizontal Scale TDS 680B 684B and 784A Test Frequency TDS 620B and 644B Test Frequency TDS 520B 540B 724A 744A Test Frequency Limits 100 mV 600 mV 6 divisions 1 ns 1 GHz...

Page 50: ...necessary to trigger a stable display At full bandwidth which for the TDS 680B TDS 684B and TDS 784A equals 1 GHz and for the TDS 520B 540B 620B TDS 644B 724A and TDS 744A equals 500 MHz frequency yo...

Page 51: ...scale setting H Enter voltage on test record H When finished checking set the horizontal SCALE back to the 50 ns setting STOP Checking each channel s bandwidth at all vertical scale settings is time c...

Page 52: ...u check a new channel H Install remove 10X attenuators between the generator leveling head and the channel input as needed to obtain the six division reference signals listed in the table i Test all c...

Page 53: ...termination and a dual input coupler See Figure 1 10 H Connect the coupler to both CH 1 and CH 2 See Figure 1 10 Sine Wave Generator Dual Input Coupler Digitizing Oscilloscope 50 Terminator Output Fig...

Page 54: ...the coupler from CH 3 to CH 4 so that CH 1 and CH 4 are driven Press CH 4 to display Be sure the vertical scale factor is kept at 1 0X See Figure 1 11 on page 1 41 Use AX1 and AX2 instead of CH3 and...

Page 55: ...r readout on screen is 100 ps for the TDS 600B or 50 ps for the TDS 500B 700A h If the channel skew is within the limits enter time on the test record and proceed to step 3 Otherwise proceed with step...

Page 56: ...e first CH1 to CH4 AX2 on the TDS 520B 620B 680B and 724A skew measurement to the second CH1 to CH4 skew measurement and divide the result by 2 Use Table 1 5 o Check against limits CHECK that the larg...

Page 57: ...strument controls a Hook up the test signal source Connect through a 50 W precision coaxial cable the time mark output of a time mark generator to CH 1 Set the output of the generator for 10 ms marker...

Page 58: ...NTAL MENU H Set horizontal modes To do this press the main menu button Time Base Press the side menu buttons Delayed Only and Delayed Runs After Main See Figure 1 13 b Measure the test signal H Set th...

Page 59: ...y Time 3 Disconnect the hookup Disconnect the cable from the generator output at the input connector of CH 1 Trigger System Checks These procedures check those characteristics that relate to the Main...

Page 60: ...E to 10 ns on the TDS 600B and 12 5 ns on the TDS 500B 700A c Hook up the test signal source Connect the output of a medium frequency leveled sine wave generator Item 25 to CH 1 Do this through a 50 W...

Page 61: ...side menu button Lower Limit Use the keypad to set the lower limit to 2 ns d Change limits until triggering stops H Press SET LEVEL TO 50 H While doing the following subparts monitor the display it w...

Page 62: ...kHz H Press the side menu button Upper Limit Use the keyboard to set the upper limit to 4 s H Press the side menu button Lower Limit Use the keypad to set the lower limit to 500 ns b Display the test...

Page 63: ...oscilloscope loses triggering is within 1 9 s to 2 1 s inclusive H Enter time on test record 4 Disconnect the hookup Disconnect the cable from the generator output at the input connector of CH 1 Equip...

Page 64: ...trigger system is within limits for Trigger level Threshold accuracy a Display the test signal H Set the vertical SCALE to 200 mV H Press VERTICAL MENU then press the main menu button Position H Set v...

Page 65: ...the side menu button for negative slope See icon at left Repeat substep b H CHECK that the Level readout in the main menu is within 9 863 V to 10 137 V inclusive for the TDS 600B or is within 9 9393...

Page 66: ...g subparts in the order listed H CHECK that the Level readout in the side menu is within 9 863 V to 10 137 V inclusive for the TDS 600B or is within 9 9393 V to 10 1147 V inclusive for the TDS 500B 70...

Page 67: ...H Press the side menu button Delayed Only then press the side menu button Delayed Triggerable H Set the horizontal SCALE for the D delayed time base to 20 ns on the TDS 600B or 25 ns on the TDS 500B 7...

Page 68: ...s the side menu button Amplitude H Press SET LEVEL TO 50 H Press CLEAR MENU H Set the test signal amplitude for about three and a half divisions on screen Now fine adjust the generator output until th...

Page 69: ...rm on both the positive and negative slopes Use the side menu to switch between trigger slopes H Enter pass fail result for main trigger on the test record H Leave the Main trigger system triggered on...

Page 70: ...ress the side menu button Main Only H Press CLEAR MENU 3 Confirm the AUX Trigger input a Display the test signal H Remove the 10X attenuator and reconnect the cable to CH 1 H Set the test signal ampli...

Page 71: ...me base then press the side menu button Delayed Triggerable H Press the side menu button Delayed Only H Set the horizontal SCALE to 500 ps for the D Delayed time base Press the side menu button Main O...

Page 72: ...the test hookup and preset the instrument controls 50 Coaxial Cables To MAIN TRIGGER OUT Calibration Generator Digitizing Oscilloscope Figure 1 20 Initial Test Hookup a Hook up test signal source 1 H...

Page 73: ...knob to center the display on screen b Measure logic levels H Press MEASURE then press the main menu button Select Measurement for Ch2 H Select high and low measurements To do this repeatedly press th...

Page 74: ...he Ch2 High readout is 1 0 volt and that the Ch2 Low readout 0 25 volts H Enter high and low voltages on test record H Press the side menu button to select the 1 MW setting H Press CLEAR MENU H CHECK...

Page 75: ...t Press the side menu button Pk Pk H Press CLEAR MENU b Check against limits H CHECK that the readout Ch2 Pk Pk is between 80 mV and 120 mV inclusive for the TDS 600B or is between 88 mV and 132 mV in...

Page 76: ...strument controls a Hook up test signal H Connect one of the 50 W cables to CH 1 See Figure 1 22 H Connect the other end of the cable just installed to the female BNC to clips adapter See Figure 1 22...

Page 77: ...purpose knob 2 Confirm that the Probe Compensator signal is within limits for frequency a Measure the frequency of the probe compensation signal H Press MEASURE then press the main menu button Select...

Page 78: ...ect the cable from CH 1 and the clips from the probe compensation terminals H Press MORE then press the main menu button Ref 1 to displayed the stored signal H Press CH 1 d Hook up the DC standard sou...

Page 79: ...U Press the side menu button AVERAGE then enter 16 using the keypad or the general purpose knob H Adjust the output of the DC calibration generator until it precisely overlaps the top upper level of t...

Page 80: ...robe Compensator Amplitude g Check against limits H Subtract the value just obtained base level from that obtained previously top level H CHECK that the difference obtained is within 495 mV to 505 mV...

Page 81: ...ults by completing the following steps a Press save recall SETUP b Press the main menu Recall Factory Setup c Press the side menu OK Confirm Factory Init d Wait for the Clock Icon to leave the screen...

Page 82: ...FT n then ENTER 3 Check Jitter vs Signal Amplitude a Set up equipment for Jitter Test See Figure 1 26 H Connect one of the rear panel composite outputs marked COMPST on the TSG121 through a 75 W cable...

Page 83: ...d be already set to 16 f Press the main menu Create Limit Test Template g Press the side menu V Limit h Use the keypad to set V Limit to 180 mdiv press 180 SHIFT m then ENTER i Press the side menu OK...

Page 84: ...the oscilloscope successfully makes 100 acquisitions If not successful the oscilloscope bell will ring When the word Run in the top left corner of the display changes to STOP the test is complete See...

Page 85: ...fter g Press the side menu RUN STOP button only h Press the main menu Mode i Press the side menu Sample j Press RUN STOP k Press VERTICAL MENU l Use the keypad to set fine scale to 300 mV press 300 SH...

Page 86: ...Test 300 mV n Use the keypad to set the fine scale to 75 mV press 75 SHIFT m then ENTER o CONFIRM that the TRIG D LED stays lighted and that the waveform on screen is stable That is it does not move...

Page 87: ...scope for 60 Hz Rejection Test H Use the keypad to set the Ch1 Fine Scale to 282 mV press 282 SHIFT m then ENTER H Press WAVEFORM OFF H Press CH2 H Press VERTICAL MENU H Use the keypad set the fine sc...

Page 88: ...or three vertical divisions of 60 Hz signal See Figure 1 32 The signal will not be triggered That is it will run free Figure 1 32 60 Hz Rejection Test Setup Signal c Check 60 Hz rejection H Use the ke...

Page 89: ...the TSG 121 through a 75 W cable and a 75 W terminator to the CH2 input See Figure 1 33 PAL Signal Source BNC T Connector 75 Terminators Digitizing Oscilloscope 50 Cable Signal Generator 75 Cable 75 C...

Page 90: ...Press the main menu Record Length H Press the side menu more until you see the appropriate menu H Press the side menu 5000 points in 100divs H Press the main menu Horiz Scale div H Use the keypad to s...

Page 91: ...tor Output Figure 1 35 Line Count Accuracy Test Hookup H Press the main menu Trigger Position H Press the side menu to Set to 50 H Press the main menu to Horiz Pos H Press the side menu to Set to 50 H...

Page 92: ...the side menu V Bars H Using the General Purpose knob place the left cursor directly over the trigger T icon H Press SELECT H Turn the General Purpose knob to adjust the right cursor for a cursor del...

Page 93: ...utton and finally Off from the side menu 7 Check the Sync Duty Cycle a Set up digitizing oscilloscope for Sync Duty Cycle Test H Press TRIGGER MENU H Press the Standard pop up to select FlexFmt Trigge...

Page 94: ...m then ENTER H Press the main menu Type pop up to select Edge H Press HORIZONTAL MENU H Press the main menu Record Length H Select the side menu 1000 points in 20div If needed first press the side men...

Page 95: ...he oscilloscope CH1 input See Figure 1 38 Pulse Generator Digitizing Oscilloscope 50 Cable 50 Terminator Output Figure 1 38 Setup for Sync Duty Cycle Test H Turn the pulse generator OUTPUT VOLTS contr...

Page 96: ...H Turn the HORIZONTAL SCALE knob to set the oscilloscope time base to 5ms div H Turn the pulse generator PERIOD variable control to adjust the period until the oscilloscope CH1 Period measurement read...

Page 97: ...urements are adjusted correctly See note above CONFIRM that the setup is correct and the oscilloscope will trigger H CONFIRM that the TRIG D LED is lighted and the waveform is stable H Disconnect the...

Page 98: ...Factory Setup and the side menu OK Confirm Factory Init Sine Wave Generator Leveling Procedure Some procedures in this manual require a sine wave generator to produce the necessary test signals If yo...

Page 99: ...a reference frequency of 10 MHz H Adjust the sine wave generator amplitude to the required number of divisions as measured by the digitizing oscilloscope 3 Record the reference level Note the reading...

Page 100: ...gitizing Oscilloscope Level Meter Power Sensor Sine Wave Generator Output Input Connect the sine wave generator to the oscilloscope and the power sensor as directed in the text Figure 1 42 Equipment S...

Page 101: ...ired new frequency H Input the correction factor for the new frequency into the level meter H Adjust the sine wave generator amplitude until the level meter again reads the value noted in step 3 The s...

Page 102: ...Performance Tests 1 88 TDS Family Performance Verification and Specifications...

Page 103: ...Specifications...

Page 104: ......

Page 105: ...44B 2 5 GS s on ea of 4 ch TDS 620B 2 5 GS s on ea of 2 ch simultaneously TDS 784A 4 GS s TDS 540B 744A 2 GS s TDS 520B 724A 1 GS s Opt 1G TDS 540B 744A 1 GS s Analog bandwidth 1 GHz on TDS 680B 684B...

Page 106: ...rameter extraction feature called measure to display the results of measurements made or the status of the instrument Assign the general purpose knob to adjust a selected parameter function More quick...

Page 107: ...are compressed to fit on the screen See Table 2 2 Table 2 2 Record Length vs Divisions per Record Samples per Division and Sec Div Sequence Record Length Divisions per Record Record Length Sample Div...

Page 108: ...erators to apply to those sources for logic pattern or state on which to trigger for mode auto or normal and for holdoff Time qualification may be selected in pattern mode Another class of logic trigg...

Page 109: ...Also you can create templates for use in limit tests On Board User Assistance Help and autoset can assist you in setting up the Digitizing Oscilloscope to make your measurements Help displays operatio...

Page 110: ...the results updated on screen as the Digitizing Oscilloscope continues to acquire waveforms An important component of the multiprocessor architecture of this Digitizing Oscilloscope is Tektronix s pr...

Page 111: ...ined are based on what is displayed on screen at the time hardcopy is invoked The hardcopies can be stamped with date and time and spooled to a queue for printing at a later time You can output screen...

Page 112: ...Specifications 2 8 TDS Family Performance Verification and Specifications...

Page 113: ...ance Selections 1 MW or 50 W Ranges Offset Volts Div Setting Offset Range Ranges Offset 1 mV div 100 mV div 101 mV div 1 V div 1 01 V div 10 V div 1 V 10 V 100 V Range Position 5 divisions Range 1 MW...

Page 114: ...100 GSamples sec TDS 784A 2 GSamples sec to 250 GSamples sec Range Seconds Division TDS 600B 0 2 ns div to 10 s div TDS 500B 724A 744A 0 5 ns div to 10 s div TDS 784A 0 2 ns div to 10 s div Record Le...

Page 115: ...ition before the clock edge and negative means a transition after the clock edge For Hold Time positive numbers mean a data transition after the clock edge and negative means a transition before the c...

Page 116: ...es with EIA TIA 574 talk only Optional on the TDS 520B and 540B Interface Centronics Centronics interface complies with Centronics interface standard C332 44 Feb 1977 REV A Interface Video VGA video o...

Page 117: ...aluminum cabinet Weight Standard Digitizing Oscilloscope 14 1 kg 31 lbs with front cover 24 0 kg 53 lbs when packaged for domestic shipment Rackmount Digitizing Oscilloscopes 14 1 kg 31 lbs plus weig...

Page 118: ...Nominal Traits 2 14 TDS Family Performance Verification and Specifications...

Page 119: ...bove this section lists only warranted characteristics A list of typical characteristics starts on page 2 23 Performance Conditions The performance limits in this specification are valid with these co...

Page 120: ...iv V div TDS 500B 700A 1 0 reading Net Offset1 Offset Accuracy 0 06 div TDS 600B 1 5 reading 0 1 div V div 0 3 mV TDS 500B 700A 1 0 reading 0 1 div 0 3 mV Accuracy Offset Volts Div Setting TDS 600B Of...

Page 121: ...pF 3 pF Input Impedance DC 50 Coupled 50 1 with VSWR 1 3 1 from DC 500 MHz 1 5 1 from 500 MHz 1 GHz Input Voltage Maximum DC 1 M AC 1 M or GND Coupled TDS 600B 400 V DC peak AC derate at 20 dB decade...

Page 122: ...ch or Pulse Width Triggering 1 ns to 1 ms 1 02 ms to 1 s 20 of setting 0 5 ns 100 ns 0 01 of Setting Input Signal Sync Amplitude for Stable Triggering NTSC and PAL modes Option 05 Video Trigger Field...

Page 123: ...el if CH 3 AUX 1 on the TDS 620B or 680B is selected as the trigger source for the main and or delayed trigger systems It is not available when a channel other than CH3 AUX 1 on the TDS 620B or 680B i...

Page 124: ...sions Emissions TDS 600B 1 2 Meets or exceeds the requirements of the following standards Vfg 243 1991 Amended per Vfg 46 1992 FCC Code of Federal Regulations 47 CFR Part 15 Subpart B Class A EN 50081...

Page 125: ...ld connected through a low impedance to both connector housings Acceptable cables are Tektronix part numbers 012 0991 00 01 02 and 03 In order to maintain the EMI performance conforming to the above r...

Page 126: ...Warranted Characteristics 2 22 TDS Family Performance Verification and Specifications...

Page 127: ...Read Out on Screen 520B 540B 724A 744A Bandwidth1 width selection is FULL TDS DS 520B 540B 724A 744A 10 mV div 100 V div 100 mV div 10 V div 50 mV div 99 5 mV div 20 mV div 49 8 mV div 10 mV div 19 9...

Page 128: ...100 V div 100 mV div 10 V div 50 mV div 99 5 mV div 20 mV div 49 8 mV div 10 mV div 19 9 mV div 500 MHz 500 MHz 500 MHz 500 MHz 450 MHz 500 MHz 500 MHz 500 MHz 500 MHz 500 MHz Accuracy Delta Time Mea...

Page 129: ...20 ps 25 ppm Reading t div 1000 Calculated Rise Time TDS 600B2 Volts Div Setting 620B 644B Rise Time 680B 684B Rise Time 10 mV div 1 V div 5 mV div 9 95 mV div 2 mV div 4 98 mV div 1 mV div 1 99 mV di...

Page 130: ...Upper 250 MHz Bandwidth Limited 250 MHz Frequency Limit Upper 20 MHz Band width Limited 20 MHz Step Response Settling Errors Volts Div Step Amplitude Settling Error 3 at Step Response Settling Errors...

Page 131: ...holdoff is 12 seconds The minimum resolution is 8 ns for settings 1 2 ms Lowest Frequency for Successful Operation of Set Level to 50 Function 30 Hz Sensitivity Edge Trigger Not DC Coupled3 Trigger S...

Page 132: ...Setting TDS 700A 8 5 ns 5 of Delta Time Setting Input Signal Sync Amplitude for Stable Triggering HDTV and FLEXFMT modes Option 05 Video Trigger All field selections 0 6 division to 4 divisions Jitte...

Page 133: ...ootnote for Sensitivity Edge Type Trigger DC Coupled in this table Stable counting of events is counting that misses no events 6 The minimum signal levels required for stable pulse width or glitch tri...

Page 134: ...Typical Characteristics 2 30 TDS Family Performance Verification and Specifications...

Page 135: ......

Page 136: ......