Troubleshooting and Error Codes
Displaying Error Codes with the Probe Degauss Autobalance Button
This section describes the error codes that the ampli
fi
ers display using the
function indicator LEDs.
When an internal error condition exists, the ampli
fi
ers may generate error codes.
To display the error codes, do the following. (See Figure 26.)
1.
Press the PROBE DEGAUSS AUTOBALANCE button.
2.
The AC and DC Coupling LEDs will
fl
ash alternately to indicate that error
codes are being displayed instead of normal conditions.
3.
The four fault indicator LEDs above the ON/STANDBY switch are used to
form a four-bit binary error code. (See Figure 27 on page 46.) The error codes
are de
fi
ned in the table. (See Table 7 on page 46.)
The PROBE OPEN LED indicates the most signi
fi
cant bit (MSB), and the
NONCOMPATIBLE PROBE TYPE LED indicates the least signi
fi
cant bit
(LSB).
4.
To continue past an error code, press any button except ON/STANDBY.
However, the degauss will fail until the internal error condition is corrected
and the degauss operation is run again.
Figure 26: Error code display
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
45