Getting Started
Connecting a Current Probe to the Ampli
fi
er
To connect a current probe to the ampli
fi
er input connector, do the following and
refer to the illustration. (See Figure 5.)
1.
To connect the probe, align the red dots.
2.
Push the probe connector in. Do not twist the connector.
3.
To disconnect the probe, pull back the collar.
4.
Pull out the connector.
Figure 5: Connecting and disconnecting a current probe to the ampli
fi
er
CAUTION.
Handle current probes with care. Do not drop a probe or subject it to
impact, or the core may crack.
Do not connect or disconnect a current probe while the probe is clamped around a
live conductor, or the probe may suffer electrical damage.
If you connect a probe to the wrong ampli
fi
er, (for example, a TCP312A to a
TCPA400), the NONCOMPATIBLE PROBE TYPE LED illuminates. Disconnect
the probe and use the correct ampli
fi
er. The TCPA400 ampli
fi
er accepts
TCP3XXA probes, but will only operate properly with TCP4XX probes.
Each current probe is calibrated before it is shipped, and should not require
further adjustment. If a probe requires adjustment, information is available in the
service manual. The adjustment procedure must be performed only by quali
fi
ed
service personnel. Contact your nearest Tektronix Service Center if you need
more assistance.
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
9
Summary of Contents for TCP300A
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...