Getting Started
Connecting the Ampli
fi
er to an Oscilloscope
You will need an oscilloscope to display the TCPA300 and TCPA400 measurement
output. To use the full dynamic range of the probe/ampli
fi
er combination, the
oscilloscope must be capable of displaying a vertical scale factor of 1 mV/div to
1 V/div.
If you are using a TEKPROBE II-compatible oscilloscope, use the
TEKPROBE-to-TEKPROBE interface cable. Otherwise, use the supplied
50
Ω
BNC cable to connect the ampli
fi
er OUTPUT connector to your oscilloscope.
The input impedance of the oscilloscope channel must be 50
Ω
, or you will
encounter slowed pulse response, increased aberrations, or incorrect DC
measurement amplitudes. If your oscilloscope provides only 1 M
Ω
inputs, you
need to attach a 50
Ω
feed-through termination between the oscilloscope input
and the BNC cable. Do not install this termination at the ampli
fi
er end of the
BNC cable.
To utilize the full bandwidth capability of the TCPA300 and TCPA400 and
attached current probe, the oscilloscope bandwidth must be approximately
fi
ve
times that of the current probe. For example, when using a TCP312A Current
Probe, the oscilloscope bandwidth must be at least 500 MHz. When using a
TCP305A Current Probe, the oscilloscope bandwidth must be at least 250 MHz.
After you have connected the ampli
fi
er to the oscilloscope, allow the equipment
to warm up to a stable temperature; usually 20 minutes is required.
Power on the Ampli
fi
er
Connect the power cord to the power input connector on the rear of the ampli
fi
er,
and then connect the power cord to your local mains supply (100 VAC to
240 VAC, 50 Hz to 400 Hz). To allow for proper ventilation, place the rear panel
of the ampli
fi
er at least 2 inches away from any obstructions. Set the ampli
fi
er on
the bottom rubber feet, and keep papers and other items away from the bottom of
the ampli
fi
er which could restrict air
fl
ow and cause overheating.
Power on the ampli
fi
er by pressing the ON/STANDBY button at the lower-left
corner of the front panel. The ampli
fi
er goes through a self-test and cycles the
front-panel LEDs.
NOTE.
The ampli
fi
er stores the power state it is in when the power cord is
unplugged. If you do not put the ampli
fi
er into STANDBY mode before unplugging
it, the ampli
fi
er will power on immediately when you plug it in again.
When you connect a probe to the ampli
fi
er, the ampli
fi
er uses detection circuitry
to indicate probe conditions such as noncompatible probe type and probe open.
8
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
Summary of Contents for TCP300A
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...