Getting Started
The TCPA300 and TCPA400 current probe ampli
fi
ers let you use one probe to
simultaneously measure AC and DC current. The ampli
fi
ers convert the sensed
current into a proportional voltage signal that you can measure directly with an
oscilloscope.
The TCPA300 and TCPA400 current probe ampli
fi
ers provide better linearity
than other current measurement systems because of a current feedback process
used with the probe. DC measurement capability and high bandwidth allow the
ampli
fi
ers to accurately represent square waves and fast-rise signals.
The TCPA300 and TCPA400 ampli
fi
ers and associated probes provide these
features:
Simultaneous DC and AC current measurements up to 750 A peak
High sensitivity
One-button autobalancing and probe degaussing
No adjustments needed to match a current probe to an individual ampli
fi
er
AC or DC coupling of signal
Direct scaling and unit readout on compatible TEKPROBE level II
oscilloscopes
System Con
fi
guration
A complete current measurement system consists of a current probe ampli
fi
er, a
compatible current probe, and an appropriate oscilloscope. (See Figure 1.)
Figure 1: Typical TCPA300/400 current measurement system
1.
50
Ω
oscilloscope input — use the TEKPROBE Interface Cable or use a
50
Ω
cable. (Add a 50
Ω
termination here if the oscilloscope only has a
high-impedance input).
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
1
Summary of Contents for TCP300A
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Page 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...