Adjustments
Adjustments
The procedures that follow describe adjustments to the probe to bring the performance within the warranted speci
fi
cations.
DC gain accuracy
Prerequisites
Complete the
DC Gain Accuracy Test
in the
Performance Veri
fi
cation
procedure. (See page 65,
Keep the test setup intact for the adjustment procedure.
CAUTION.
To avoid ESD damage to the probe, use an antistatic wrist strap and work at a static-approved workstation when
handling the probe.
72
20 A Current Probes Instruction Manual