Preface

products, you will need to either press a rear-panel OFF switch or remove the
power source from the instrument.

Remove.

This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product Service
Manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot

be recovered by other means or technology. This is typically used when the
device will be moved (temporarily or permanently) from a secured area to a
non-secured area.

Scrub.

This is equivalent to sanitize.

User Accessible.

User is able to directly retrieve the memory device contents.

User-modifiable.

The user can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

User-modifiable.

The user can write to the memory device during normal

instrument operation, using the instrument interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

Device terms

User data.

Describes the type of information stored in the device. Refers to

waveforms or other measurement data representing signals connected to the
instrument by users.

User settings.

Describes the type of information stored in the device. Refers

to instrument settings that can be changed by the user.

Both.

Describes the type of information stored in the device. It means that

both user data and user settings are stored in the device.

None.

Describes the type of information stored in the device. It means that

neither user data nor user settings are stored in the device.

Directly.

Describes how data is modified. It means that the user can modify

the data.

Indirectly.

Describes how data is modified. It means that the instrument

system resources modify the data and that the user cannot modify the data.

iv

TTR500 Series and TCAL500 Declassification and Security Instructions

Summary of Contents for TCAL500

Page 1: ...TTR500 Series Vector Network Analyzers and TCAL500 Calibration Kit Declassification and Security ZZZ Instructions xx www tek com 077 1256 01...

Page 2: ...ssued and pending Information in this publication supersedes that in all previously published material Speci cations and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Page 3: ...Table of Contents Preface iii Memory devices 1 Clear or sanitize a non functioning instrument 3 TTR500 Series and TCAL500 Declassi cation and Security Instructions i...

Page 4: ...Table of Contents ii TTR500 Series and TCAL500 Declassi cation and Security Instructions...

Page 5: ...documents are required to perform the procedures in this document Terms used in this document Clear This removes data on media memory before reusing it in a secured area All reusable memory is cleare...

Page 6: ...uring normal instrument operation using the instrument interface or remote control Volatile memory Data is lost when the instrument is powered off Device terms User data Describes the type of informat...

Page 7: ...ds Disconnect power for at least 20 seconds SRAM 128KB Not applicable Not applicable No Not applicable Not applicable RFIC No Disconnect power for at least 20 seconds Disconnect power for at least 20...

Page 8: ...GB Software installer documentation None unless written by user Host OS contains option to write files to disk Standard Windows file copy Embedded USB memory module Yes Disable USB storage in VectorVu...

Page 9: ...device embedded in the VNA No user data is stored on the instrument when it is disconnected from a power supply To clear or sanitize the instrument 1 Enable USB storage in activation manager in Vector...