5Ć8
# $
! !
One time markgenerator (item 9), one precision
coaxial cable (item 5), and one 50
W
termination (item 3).
!
CRT adjustment procedure.
NOTE
To enable factory calibration on oscilloscopes with firmware version
2.10
or above, remove jumper J205 (CAL-DIS) on the CPU board.
(SeeFigure5Ć17 on page5Ć23 for thelocation of J205.) If you do
not remove this jumper, the message See manual to enable FactoĆ
ry Cal" appears when you select the factory calibration routine.
Firmware versions below
2.10
do not havethecalibration lockout
featureand thefactory calibration routinecan still executewith J205
installed.
1. Disconnect all signal inputs from the oscilloscope.
2. Press the
button to display the Utility menu.
CAUTION
To avoid replacing the calibration constants with erroneous values,
besureto completeall steps of this procedureafter initiating the
calibration routine.
3. Select
from the main menu.
4. Select
# $
, initiating the calibration routine.
This routine takes about 20 seconds to complete its initial calibrations.
Once complete, 2 dots are displayed and the message
" !
" appears. The displayed dots are used for the next
horizontal calibration steps.
NOTE
Thefollowing steps useboth theGeneral PurposeKnob and the
TOGGLE
button to make adjustments. The General Purpose Knob
performs two functions: to adjust horizontal gain and centering. The
TOGGLE
button selects the function of the General Purpose Knob.
Summary of Contents for TAS 465
Page 8: ...vi...
Page 12: ...x...
Page 16: ...xiv...
Page 17: ...xv...
Page 18: ...xvi...
Page 24: ...1 4...
Page 38: ...1 18...
Page 43: ...2 5...
Page 49: ...0 3 3 1...
Page 66: ...4 12...
Page 84: ...4 30...
Page 86: ...5 2 Display Driver board Analog board...
Page 103: ...5 19 D D...
Page 104: ...5 20 R321 R322 R323 R112 R154 R111 R140 R141 C121 C122...
Page 105: ...5 21 Ch2 10X Ch2 100X Ch1 10X Ch1 100X...
Page 106: ...5 22 R43...
Page 107: ...5 23 J205 CAL DIS...
Page 108: ...5 24...
Page 109: ...6 1 H H H H H H...
Page 112: ...6 4...
Page 116: ...6 8...
Page 123: ...6 15 1 Unplug 2 Rotate 3 Pull Line Fuse...
Page 125: ...6 17 Front Foot Rotate and remove Mounting Screws 4 Mounting Screw...
Page 129: ...6 21 Spudger EMI Gaskets 4 H H H...
Page 136: ...6 28 Display Driver Board A3 Mounting Screw High Voltage Cover Mounting Screw H H H...
Page 137: ...6 29 CRT Implosion Shield Spudger Slot In CRT Bracket Implosion Shield Guides...
Page 139: ...6 31 Fan Mount Flex locks Fan...
Page 148: ...6 40...
Page 160: ...6 52...
Page 168: ...8 6...
Page 169: ...8 7...
Page 186: ...8 24...
Page 191: ...8 29...
Page 195: ...3 1 1 0 0 1 2 1 1 0 0 1 2 2 10 10 0 10 1 0 0...
Page 197: ......
Page 200: ......
Page 202: ......
Page 206: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ...10 1 0 0 0 0 0 H H 0 H H 0 0 0 0 0 0 1 0 0 1 1...
Page 258: ...10 2...
Page 261: ...10 5...
Page 263: ...10 7...
Page 265: ...10 9...