
SPG8000A User Manual
119
K
T
parameter
Display name: KT (T
Freq.)
Units: cycles per second (c/s)
The K
T
(T Freq.) parameter, if used alone, does not create a “pattern.” Rather, the
entire screen will cycle (sinusoidally) between white and Black. The effect will
be
similar on either line or field display of a waveform
monitor.
Typically, a small K
T
coefficient is set to add apparent motion to a standard circle,
sine wave, or sweep pattern. To see the picture effects of this parameter, select
any of the standard zone plates and vary K
T
. Note that the K
T
coefficient may be
positive or negative; the sign will determine the direction of the pattern’s motion.
K
Y
parameter
Display name: KY (V
Freq.)
Units: cycles per active picture height (c/aph)
The K
Y
(V Freq.) parameter sets the initial
vertical
frequency of the pattern (it is
analogous to K
X
rotated 90° clockwise).
To
see the effects of this parameter, set it
to a low number and adjust a waveform monitor to view one or two fields of the
i
nstrument SDI output. The variation in the (dc) voltage level from line to line
will give the field waveform the appearance of a sampled sine wave, especially at
a
K
Y
of
around
5 to 10
c/aph
(the
sinusoid
can
become
difficult
to discern
at
higher
frequencies). Note that in the Y (vertical) parameters, the c/aph units translate
directly into visible cycles down the
frame.
The Vertical Sine zone plate is a “K
Y
-
only” zone plate. (See Figure 35.)
Figure 35: A “K
Y
-only” zone plate
K
T
2
parameter
Display name: KT^2 (T
Sweep)
Units: cycles per second per second (c/s^2)
The K
T
2
(T Sweep) parameter sets the change in temporal frequency over time. If K
T
is initially 0.0, setting K
T
2
to 1.0 will result in an apparent K
T
of 10.0 after ten
seconds, 20.0 after twenty seconds, and so forth. The temporal frequency will
increase to maximum, return to 0.0, and repeat the cycle as long as the value of K
T
2
remains greater than zero.
Summary of Contents for SPG8000A
Page 53: ...SPG8000A User Manual 39 Figure 7 Removing a Power Supply module...
Page 103: ...SPG8000A User Manual 89 Figure 22 SDI Overlay submenu...
Page 182: ...168 SPG8000A User Manual Figure 61 AES button menu diagram...
Page 190: ...176 SPG8000A User Manual Figure 73 EMBED button menu diagram...
Page 194: ...180 SPG8000A User Manual Figure 77 DOLBY E CONFIG submenu diagram part 2...
Page 197: ...SPG8000A User Manual 183 Figure 81 REF button menu diagram part 2...
Page 200: ...186 SPG8000A User Manual Figure 84 SDI button menu diagram part 2...
Page 211: ...SPG8000A User Manual 197 Figure 96 SDI ANC PAYLOAD submenu diagram part 2...
Page 213: ...SPG8000A User Manual 199 Figure 97 STATUS button menu diagram part 1...
Page 215: ...SPG8000A User Manual 201 Figure 99 SYSTEM button menu diagram part 1...
Page 216: ...202 SPG8000A User Manual Figure 100 SYSTEM button menu diagram part 2...
Page 225: ...SPG8000A User Manual 211 Figure 107 SNMP TRAPS submenu diagram part 2...
Page 240: ...226 SPG8000A User Manual Figure 123 PTP primary master menu diagram part 1...
Page 241: ...SPG8000A User Manual 227 Figure 124 PTP primary master menu diagram part 2...
Page 243: ...SPG8000A User Manual 229 Figure 126 PTP primary slave menu diagram part 2...