Performance Verification
SD-44 Instruction Manual
13
Table 2: Test equipment (cont.)
Description
Example product
Minimum requirements
Sampling oscilloscope with
sampling head
11K (1180X, CSA80xX with
SD-32 sampling head)
PC with GPIB port and
printer
printer output of sampled wave-
forms
Adjustable single-mode
optical attenuator
4 decades, 9
m
m core fiber, FC-
style connectors
Tektronix OA5002
Digital voltmeter
4 1/2 digit
Keithley 2000 or
Tektronix DMM916
50
W
termination
±
1%
011-0049-01
BNC-to-banana adapter
BNC female to dual banana
103-0090-00
Optical cable
FC-FC multimode, 9
m
m, 2 meters
174-1910-00
Inline optical adapter
FC female to FC female
131-5039-00
1
CW and modulated mode available: modulation with OFF level at or below 0.1
m
W,
optical falltime < 1
m
s
The O/E converter under test and the test equipment and should be warmed up
for 20 minutes at an ambient temperature between 20 and 30
_
C.
Output Zero
1. Attach the output of the SD-44 to the voltmeter inputs with a 50
W
termina-
tion and BNC-to-banana adapter.
2. Install the optical dust cover on the input of the SD-44.
3. Check that output voltage is
≤
±
1 mV. Record the result on the test record.
DC Conversion Gain
NOTE
. Make sure that the optical connector ends of both the fiber from the
optical attenuator output and the SD-44 under test input fiber are well cleaned
before performing this step. See to the cleaning instructions on page 7.
1. Connect the 1310 nm laser source to attenuator input.