![Tektronix SD-42 User Manual Download Page 26](http://html1.mh-extra.com/html/tektronix/sd-42/sd-42_user-manual_1078303026.webp)
Performance Check
SD-42 & SD-46 User Manual
3–3
Optical Power Monitor Check
The Optical Power Monitor function of the SD-42/SD-46 does not use any of the
oscilloscope functions. However the oscilloscope must be switched on and the
electrical output of the SD-42/SD-46 must be terminated, preferably using the
high-performance U-link connected to an adjacent SD-24/26.
WARNING.
Avoid eye exposure to the output of open-ended fibers/connectors.
Ensure all fibers are terminated prior to turning on an optical source.
1. Connect the equipment as shown in Figure 3–1.
OA5002
Optical Cable
Laser
DVM
CSA 803
SD-42/SD-46
SD-22/SD-26
Optical
Input
Rigid
U-link
+
—
Out
In
Out
Figure 3–1: Setup to Check Optical Power Monitor Operation
2. Switch on the CW laser source.
3. Set the digital voltmeter as follows:
DC Volts
Autoranging (or 200 mV)
4. On the SD-42/SD-46, using SELECT RANGE, set the range to 1 V/mW.
5. Adjust the optical attenuator until the digital multimeter reads 100 mV.
There is now 100
m
W of power incident on the SD-42/SD-46 input.
6. Increase the setting of the optical attenuator by 30 dB.
Summary of Contents for SD-42
Page 2: ...User Manual SD 42 SD 46 Optical to Electrical Converter Heads 070 8671 01...
Page 5: ......
Page 10: ...Getting Started...
Page 11: ......
Page 15: ...Getting Started 1 4 SD 42 SD 46 User Manual...
Page 16: ...Operating Basics...
Page 17: ......
Page 21: ...Operating Basics 2 4 SD 42 SD 46 User Manual...
Page 22: ...Reference...
Page 23: ......
Page 29: ...Performance Check 3 6 SD 42 SD 46 User Manual...
Page 30: ...Appendices...
Page 31: ......
Page 33: ...Appendix A Accessories A 2 SD 42 SD 46 User Manual...
Page 44: ...Index...
Page 45: ......