Reference > Connecting Signals
Connecting a Signal Using a TekConnect Probe
Connecting a Signal Using a TekConnect Probe
You can connect a signal to the spectrum analyzer using a TekConnect Probe by using an RTPA2A
Real-Time Spectrum Analyzer TekConnect Probe Adapter. The RTPA2A enables you to use Tektronix
P7000 Series active and differential probes to connect to signals. See the RTPA2A Instruction Manual for
instructions on how to connect the probe adapter and a probe to the spectrum analyzer.
To use the RTPA2A probe adapter with the RSA6100A Series instruments, you may need install a device
driver for the RTPA2A probe adapter.
Installing the RTPA2A Probe Adapter Device Driver
When the RTPA2A probe adapter is connected to the instrument’s USB port for the
fi
rst time, the Windows
XP Add Hardware wizard may automatically start. If the wizard does start, use the procedure below to
install the device driver for the new hardware. (If the wizard does not start, the probe adapter device
driver is already installed.)
To properly install the probe adapter device driver:
1.
Select the option for Windows to install from a speci
fi
c location (then select
Next
).
2.
Only search for the driver in
C :\Program Files\Tektronix\RSA6100A\RSA\ProbeAdapter
.
3.
If Windows XP warns about Windows Logo testing, select
Continue Anyway
.
4.
Select
Finish
after the Found New Hardware wizard completes.
Connecting External Trigger Signals
External trigger signals are connected to the spectrum analyzer using the
Trig In
connector on the front
panel. There is also an additional trigger input on the rear panel, the Trig 2 In.
To connect a trigger signal to the spectrum analyzer, connect the signal to the
Trig In
connector on the
front panel, or the
Trig 2 In
connector on the rear panel.
Table 6: External trigger signal input requirements
Input
Level
Impedance
Trig In
-2.5 V to +2.5 V
50
Ω
or 5,000
Ω
Trig 2 ln
+5 V (TTL)
10,000
Ω
xxx
Enabling External Trigger
To use an external trigger signal, select either
Trig In (front)
or
Trig 2 In (rear)
in the
Source
control on
the Trigger control panel’s
RSA6100A Series Real-Time Spectrum Analyzers Help
345