Controlling the Acquisition of Data > Using Triggers to
Capture Just What You Want > Triggering
Actions Tab (Triggering)
Actions Tab (Triggering)
The Actions tab sets parameters for the Save on trigger function. The Save on trigger function allows you
to save an acquisition data
fi
le and/or a screen capture when a trigger event occurs.
Trigger events are de
fi
ned on the
and
tabs.
Action tab settings
Setting
Description
Save acq data on Trigger
Enables the Save on Trigger function. The list box to the right of the
Save acq
data on Trigger
check box speci
fi
es the format of the saved
fi
le. Acquisition
data
fi
les can be saved in TIQ (native) format, CSV format, and MAT (Matlab)
format. For descriptions of these
fi
le formats, see
.
When TIQ is the selected
fi
le type, a
fi
le is saved after a complete acquisition,
which can include multiple frames. When CSV or MAT is the selected
fi
le type, a
separate
fi
le is saved for each frame.
Save picture on Trigger
Enables saving a picture of the display when a trigger event occurs. Files can be
saved in PNG, JPG, and BMP format. For descriptions of these
fi
le formats, see
Data, Settings, and Picture File Formats (see page 335)
Max total
fi
les saved per Run
Speci
fi
es the number of times a Save on Trigger action stores a
fi
le. After this
limit is reached, no more
fi
les are saved. The instrument will continue to run, but
no additional
fi
les are saved when a trigger event occurs.
Keep in mind when setting this number that picture
fi
les are counted as part of the
total number of
fi
les. For example, if you set Max total
fi
les to 100, the instrument
will save 100 acquisitions if only acquisitions are saved or only pictures are
saved. But, if both acquisitions and pictures are saved, then 50 acquisitions
and 50 pictures will be saved.
xxx
RSA6100A Series Real-Time Spectrum Analyzers Help
333
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...