RF Measurements > Common Controls for Settling Time
Displays
Settling Time Displays Shared Measurement
Settings
Settling Time Displays Shared Measurement Settings
The control panel tabs in this section are shared by the Frequency and Phase Settling Time displays. These
control panel tabs are not shared with other RF Measurement displays. With few exceptions, each display
shares values for each of the controls on these tabs. For example, Vertical Scale values are independent for
each of the displays (degrees for Phase Settling Time and Hertz for Frequency Settling Time).
Common controls for settling time measurement displays
Settings tab
Description
Speci
fi
es the parameters that characterize the settling time measurement.
Speci
fi
es the Measurement length and minimum settled duration for the settling time
measurement.
Speci
fi
es the parameters used for Mask testing.
Specify which traces to show and how they are computed.
Speci
fi
es the vertical and horizontal scale settings.
Speci
fi
es display parameters of the Settling Time displays.
xxx
De
fi
ne Tab for Settling Time Displays
The De
fi
ne tab for the Frequency Settling Time measurement controls several parameters that affect
how the measurement is made.
Meas Freq
Select the frequency of the signal to be measured. Measurement Frequency can be selected as the target
frequency (see
Target Reference
).
NOTE.
If Lock Center Frequency of Spectrum Display to Measurement Frequency is unchecked, then
Measurement Frequency is independent of Center Frequency) Range: Entire input frequency of the
instrument
190
RSA6100A Series Real-Time Spectrum Analyzers Help