RF Measurements > Occupied BW & x dB BW
Parameters Tab
Parameters Tab
The Parameters tab enables you to specify parameters that control the Occupied BW measurement.
Setting
Description
Occupied BW % Power
Speci
fi
es the proportion of power within the occupied bandwidth (referenced against the
total power in the measurement bandwidth).
x dB level
x dB level de
fi
nes the x dB BW level search threshold.
Measurement BW
Speci
fi
es the frequency range used by the measurement.
Max Hold spectral data
Enables the Max Hold function.
Average results
Enables/disables results averaging. Note that this is not an averaging of the trace, but an
averaging of the results.
Count
Speci
fi
es the number of results averaged to calculate the Occupied BW. Range: 2 to
10,000.
xxx
x dB Level
The x dB level determines the x dB BW. The instrument analyzes the spectrum trace to locate the
frequencies at which the level is x dB down from the peak level calculated over the measurement
bandwidth. The frequency difference between the upper and lower crossing thresholds is the x dB BW.
Range: -80.0 to -1.0 dB; Resolution: 0.1%; Inc/dec small: 0.1%, large: 1%; Default: -26 dB
Max Hold Spectral Data
Max Hold displays the maximum value in the acquisition record for each display point. Each new trace
display point is compared to the previous maximum value and the greater value is retained for display
and subsequent comparisons.
152
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...