![Tektronix RSA6100A Series Help Manual Download Page 122](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169122.webp)
Analog Modulation > AM
AM Settings
Item
Element
Description
1
Check mark indicator
The check mark indicator in the upper, left-hand corner of the display shows
when the AM display is the optimized display.
2
Top of Graph
Sets the %AM indicated at the top of the graph by increasing or decreasing
the vertical scale. Changing the top value affects the bottom of graph value
because the graph scales about vertical center. Also, note that the top of
graph setting interacts with the internal vertical scale setting (which is not
user settable).
3
Position
Speci
fi
es the %AM shown at the center of the graph display.
4
Bottom Readout
Displays the value of the modulation factor shown at the bottom of graph.
5
Measurement readouts
Displays numeric values for the +AM (positive modulation factor), –AM
(negative modulation factor), and Total AM.
6
Position
Speci
fi
es the horizontal position of the trace on the graph display.
7
Scale
Adjusts the horizontal range of the graph. By decreasing the scale (time
across the entire graph), the graph essentially becomes a window that you
can move over the trace by adjusting the position.
8
Clear
Erases the trace from the graph.
9
Trace Detection readout
Displays the Settings > Trace > Detection setting.
xxx
Changing the AM Settings (see page 110)
AM Settings
Menu Bar: Setup > Settings
Application Toolbar: Settings
The AM Settings control panel provides access to settings that control parameters of the AM Display.
Setting
Description
Sets the Carrier Amplitude Detection method. You can choose either Average or Median.
Sets Trace display parameters.
Sets vertical and horizontal scale and position parameters.
Speci
fi
es whether or not certain display elements are shown. Speci
fi
es the maximum
number of points shown in the display graph.
xxx
110
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...