Warranted characteristics
When the instrument is installed according to the instructions in this document, the rackmounted instrument meets all warranted
requirements listed in the instrument specification. Instruments mounted using methods other than those described in these
instructions might not meet their warranted requirements.
For tables of the warranted characteristics, see Specifications in the specification and performance verification manual that
applies to your instrument model.
CAUTION. The instrument might be damaged due to overheating. When installing multiple instruments in an instrument rack, the
ambient temperature may go up. You assume the responsibility to provide adequate cooling to meet the ambient temperature
requirements listed in the specifications.
The ambient temperature inside the instrument rack will vary depending on the location of the instrument within the instrument
rack. Tektronix recommends that you measure the ambient temperature in the desired rack location before you install the
instrument to ensure the operating temperature does not exceed the rated ambient temperature limit. If necessary, refer to the
Environmental Specifications in your product documentation for the operating temperature limits.
Clearance requirements
The rack in which the individual instrument is mounted must provide the following clearance requirements:
■
At least 218 mm (8.6 in) of vertical space
■
A minimum width of 488 mm (19.2 in) between the left- and right-front rails in the rack. If possible, consider allowing at least
50.8 mm (2.0 in) on the right side and rear of the instrument for adequate airflow.
■
A minimum inside depth of at least 384 mm (15.1 in) depth (from rack mounting ear to back of instrument)
Airflow enters from the back of the instrument and exits on the right of the instrument when viewed from the front.
CAUTION. Adhering to these requirements mounts the rack-adapted instrument with enough clearance for air circulation and
accommodation of the power cord and mounting hardware. Failure to provide these clearances can result in overheating and can
cause the instrument to operate improperly or fail.
Kit description
RM5 Rackmount Kit
3
Summary of Contents for RM5
Page 4: ...Table of Contents ii RM5 Rackmount Kit ...
Page 6: ...Important safety information iv RM5 Rackmount Kit ...
Page 10: ...Kit description 4 RM5 Rackmount Kit ...
Page 17: ...Dimensions Front panel Figure 8 5 Series MSO front panel RM5 Rackmount Kit 11 ...
Page 18: ...Figure 9 6 Series MSO front panel Dimensions 12 RM5 Rackmount Kit ...
Page 19: ...Rear panel Figure 10 5 Series MSO and 6 Series MSO rear panel Dimensions RM5 Rackmount Kit 13 ...