background image

Copyright

E

Tektronix. All rights reserved.

P80318
Handheld Differential
Impedance Probe
Instructions

www.tektronix.com

Safety Summary

To avoid potential hazards, use this product only
as specified.

To avoid fire or personal injury

Ground the product.

This product is indirectly

grounded through the grounding conductor of
the mainframe power cord. To avoid electric
shock, the grounding conductor must be con-
nected to earth ground. Before making connec-
tions to the input or output terminals of the
product, ensure that the product is properly
grounded.

Connect and disconnect properly.

Do not connect

or disconnect probes or test leads while they are
connected to a voltage source.

Observe all terminal ratings.

To avoid fire or shock

hazard, observe all ratings and markings on the
product. Consult the host instrument documenta-
tion for further ratings information before mak-
ing connections to the host instrument.

Do not apply a potential to any terminal,
including the common terminal, that exceeds the
maximum rating of that terminal.

Do not operate with suspected failures.

If you

suspect there is damage to this product, have it
inspected by qualified service personnel.

Do not operate in wet/damp conditions.

Do not operate in an explosive atmosphere.

Keep product surfaces clean and dry.

Safety symbols on the product

CAUTION

Refer to Product Manual

Product End-of-Life Handling.

Observe the fol-

lowing guidelines when recycling an instrument
or component:

Equipment Recycling.

This product complies

with the European Union’s requirements accord-
ing to Directive 2002/96/EC on waste electrical
and electronic equipment (WEEE). For informa-
tion about recycling options, check the Support/
Service section of the Tektronix Web site
(www.tektronix.com).

Restriction of Hazardous Substances.

This

product has been classified as Monitoring and
Control equipment, and is outside the scope of
the 2002/95/EC RoHS Directive. This product is
known to contain lead and hexavalent chromium.

Warranty Information

For warranty information, go to
www.tektronix.com, click Support, and then
select Look Up Tektronix Warranty.

Probe tips and the ground wires are consumable
items and are not covered by the warranty.

Standard Accessories

The P80318 probe includes the following items:

H

2 one-meter SMA cables, each
with a parallel control line

H

3 ground wires (1 installed, 2 spares)

H

Adjustment tool

H

Vinyl probe tip protection cap

H

2 strap sets for fastening together the cables.

Product Description

The P80318 High Performance 100 Ohm
Differential Impedance Hand Probe provides
users a differential hand probe with adjustable
spacing tips, EOS/ESD protection (with the
Tektronix 80A02 EOS/ESD Protection Module),
and a unique double spring mechanism that
provides physical feedback to optimize probing
contact pressure. The probe is suitable for
probing differential runs on circuit boards, IC
package pins, and other devices that have a
0.5 mm to 4 mm pitch.

The P80318 probe is designed to be used with
the Tektronix 80A02 EOS/ESD Protection
Module. Together, they protect sensitive
instrument-input stages, such as the sampling
bridge of Tektronix electrical sampling modules,
from electro-overstress (EOS) and electrostatic
discharge (ESD) damage due to static charges
stored in the device under test (DUT).

Connecting the Probe

80A02 Module

Control

Input

Output

Sampling

module

Input

P80318 probe

DUT

50

Ω

termination

Repeat for second cable

Connector Care

Use extra care when attaching or removing the
cable SMA connectors from the probe. Turn only
the nut, not the cable. When attaching the cable,
align the connectors carefully before turning the
nut. Use light finger pressure to make this initial
connection. Then tighten the nut lightly with a
wrench. Ideally, tighten the SMA connectors to
56 N-cm (5 in-lb) with a torque wrench.

Adjusting Probe Tip Gap

Adjustment hole

Adjustment tool

Adjustment

tension

Probing Guidelines

H

The P80318 uses a unique double spring
mechanism that provides physical feedback
to optimize probing contact pressure. When
applying pressure after the initial contact to
the DUT, the first spring engages and the
probe tips retract with the first spring pres-
sure. As you apply more pressure, the second
spring engages, increasing the probe com-
pression resistance. This sudden increase in
stiffness is the correct operation zone. In-

*P071177501*

071-1775-01

Reviews: