![Tektronix P7313SMA Technical Reference Download Page 37](http://html1.mh-extra.com/html/tektronix/p7313sma/p7313sma_technical-reference_1079068037.webp)
Speci
fi
cations
Speci
fi
cations
The speci
fi
cations in the following tables apply to a P7313SMA probe installed on a TDS6804 oscilloscope. The probe must
have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits described. (See
Table 4.) Speci
fi
cations for the P7313SMA differential probe fall into three categories: warranted, typical, and nominal
characteristics.
Warranted Characteristics
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. (See
Table 4.) Warranted characteristics that have checks in the
Performance Veri
fi
cation
section are marked with the
symbol.
Table 4: Warranted electrical characteristics
Characteristic
Description
Differential rise time, 10-90% (probe only) (Main output)
≤
40 ps, +20 °C to +30 °C (+68 °F to +86 °F),
100 mV differential step in 2.5X attenuation
500 mV differential step in 12.5X attenuation
Differential signal range
0.800 Vp-p (2.5X attenuation)
3.6 Vp-p (12.5X attenuation)
DC gain (Main output)
0.40 ±2.0% (corresponds to 2.5 X attenuation)
0.08 ±2.0% (corresponds to 12.5 X attenuation)
Termination voltage range
+3.6 V /-2.5 V
Termination voltage accuracy
(EXT mode)
(INT mode)
(AUTO mode)
±(0.2% x V
T
+2 mV) over a +3.6 V/–2.5 volt V
T
range
±(0.3% x V
T
+ 2 mV) over a +3.6 V/–2.5 volt V
T
range
±(2.5% x V
CM
+ 20 mV) over a +3.6 V/–2.5 volt V
CM
range
Linearity
±1% or less of dynamic range
Output offset voltage (Main output) V
CM
= 0 V, V
DM
=
0 V, V
T
= 0 V
±2.5 mV +20 °C to +30 °C (+68 °F to +86 °F)
Differential-mode input resistance
100
Ω
±2%
Maximum nondestructive input voltage V
T
= 0 V, applied
<5 minutes
±5 V (DC + peak AC) on either SMA input
Maximum nondestructive external termination input voltage
±15 VDC
Temperature
Operating: 0 to +40 °C (+32 to +104 °F)
Nonoperating: -55 to +75 °C (-131 to +167 °F)
Humidity
Operating: 0-90% RH, tested at +30 to +40 °C (+68 to
+104 °F)
Nonoperating: 0-90% RH, tested at +30 to +60 °C (+68 to
+140 °F)
P7313SMA Technical Reference
25
Summary of Contents for P7313SMA
Page 1: ...P7313SMA 13 GHz Differential Probe Technical Reference P071196800 071 1968 00...
Page 2: ......
Page 3: ...P7313SMA 13 GHz Differential Probe Technical Reference www tektronix com 071 1968 00...
Page 6: ......
Page 8: ...Table of Contents ii P7313SMA Technical Reference...
Page 12: ...Preface vi P7313SMA Technical Reference...
Page 34: ...Operating Basics 22 P7313SMA Technical Reference...