![Tektronix P7240 User Manual Download Page 33](http://html1.mh-extra.com/html/tektronix/p7240/p7240_user-manual_1078107033.webp)
P7240 4 GHz 5X Active Probe User Manual
23
Appendix A: Specifications
These specifications apply to a P7240 probe when used with a
TDS7404 oscilloscope.
The probe and oscilloscope must first be allowed to warm up for
20 minutes before measurements are taken.
Specifications
CAUTION.
To prevent damage to the probe or circuit under test, do
not apply voltages beyond the nondestructive input voltage range to
the probe.
Table 3: Warranted electrical specifications
DC Attenuation Accuracy (probe only)
0.2
±
2% (excludes offset error)
Output Zero
±
10 mV or less at output of probe
Rise Time (probe only)
≤
120 ps
Table 4: Typical electrical characteristics
Bandwidth, (probe only)
4 GHz (See Figure 10)
Linear Input Dynamic Range
--2.0 V to +2.0 V
(Equivalent to --0.4 V to +0.4 V
at the output of the probe.)
Linearity
ᐔ
0.1% over a dynamic range of
--1.75 V to + 1.75 V
ᐔ
1.0% over a dynamic range of
--2.00 V to + 2.00 V
Nondestructive Input Voltage Range
--30 V to +30 V (DC + peak AC)