P7225 2.5 GHz 10X Active Probe User Manual
35
Appendix B: User Service
This section explains the maintenance for the P7225 probe.
Inspection and Cleaning
CAUTION.
To prevent damage to probe materials, do not use
chemicals that contain benzine, benzene, toluene, xylene, acetone, or
similar solvents.
Do not immerse the probe or use abrasive cleaners.
Dirt may be removed with a soft cloth dampened with a 75%
isopropyl alcohol and water solution.
Calibration
If you need to do a performance verification or make internal
adjustments to your probe, refer to the
P7225 2.5 GHz Active Probe
Service Manual.
The manual is a printable pdf file, and is available
on both the Tektronix website and the Optional Applications CD.
The Optional Applications CD is included with Tektronix oscillo-
scopes and signal analyzers featuring the TekConnect interface.
Replacement Parts
Refer to the
Replaceable Parts
section for a list of customer
replacement parts. Due to the sophisticated design of the P7225
probe, there are no user replaceable parts within the probe.