Midband Accuracy
This test checks the midband accuracy of the probe. If the measurements are out of the speci
fi
ed limits in the test record, refer to
the
Adjustments
section. (See page 49.)
Setup
1.
Connect a BNC-to-Dual Banana adapter to
the digital multimeter (DMM) input.
2.
Connect the Leveled Sinewave Output
of the calibration generator to a BNC “T”
connector and then connect the other side
of the BNC “T” to the DMM input.
3.
Connect the middle branch of the BNC “T”
connector to the calibration
fi
xture.
P6021A Current Probe Instruction Manual
40