Adjustments
Adjustments
The procedures that follow describe adjustments to the probe to optimize the performance within the warranted speci
fi
cations.
Midband accuracy
Aberrations
Prerequisites
Complete the
Performance Veri
fi
cation
procedures.
CAUTION.
To avoid ESD damage to the probe, use an antistatic wrist strap and work at a static-approved workstation when
handling the probe.
Equipment Required
Refer to the Performance Veri
fi
cation procedures for the equipment required. You also need a #2
fl
at-bladed screwdriver and an
insulated,
fl
at-bladed adjustment tool. The tool is available from Tektronix; order part number 003-1433-00.
49
P6021A Current Probe Instruction Manual