Appendix G: Product Verification
OA 5000 Series User Manual
A ---23
Step 1:
Push the
MODE
button until ATT is backlit and -REF is not
backlit (these LEDs are green);
H
The
MODE
button itself should now be backlit (this LED is red).
Step 2:
Push the
MIN ATT
button; this will cause the instrument to
adjust to zero dB (a few seconds may be needed to complete the ad-
justment if the previous setting was non-zero).
H
The
MIN ATT
button should itself be backlit
H
0.00
should be displayed in the ATTENUATION(dB) display
Step 3:
Push the
SET REF
button.
H
The
SET REF
button should now be backlit; the
MODE
button
should not be backlit.
H
Check that the WAVELENGTH(nm) display is showing “
SEt
”.
Step 4:
Adjust the numerical value displayed in the ATTENUATION(dB)
display by manipulating the
COARSE
and
FINE
knobs; the coarse knob
should increment or decrement the value by 1 dB per detent; the fine
knob should increment or decrement the value by 0.01 dB.
H
Set the display to read “
-- 10.00
”
H
Push the
SET REF
button again; the button will then be unlit and the
MODE
should now be lit with the ATT only backlit.
Step 5:
Push the
MODE
once again.
H
Check to see that “ATT” and “--- REF” are both backlit
H
The ATTENUATION(dB) display should now read “
10.00
” (This is
because ATT=0.00;REF= --- 10.00; hence, the result
ATT --- REF=0 --- ( --- 10.00)=10.00)
H
Pushing the
MODE
button repeatedly will toggle the “--- REF” back-
lighting on and off and the displayed attenuation will toggle between
“
0.00
” and “
10.00
”
H
Set the
MODE
so that the “ATT” only is backlit before going to the
next step.
Step 6:
Push the
WAVELENGTH
button once.
H
The ATTENUATION(dB) display should read “
SEt
” and the
WAVELENGTH
button will be backlit.
H
The
COARSE
and
FINE
knobs increment or decrement the value
displayed in the WAVELENGTH(nm) display; the coarse knob should
increment or decrement the wavelength value by 10 nm; the fine
knob should increment or decrement the value by 1 nm.
Display Modes
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for OA 5000 Series
Page 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 21: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 29: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 93: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 131: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...