Connect the probe tip to the 1 kHz source, and the ground clip to the ground. For best results, remove any probe tip
accessories and hold the probe tip directly onto the 1 kHz connector.
NOTE. Connect only one probe at a time to the PROBE COMP terminals.
3. Turn off all channels.
4. Turn on the channel to which the probe is connected.
5. Select File > Autoset from the menu bar. The screen displays a square wave.
6. Double-tap the badge of the channel that you want to compensate.
7. Tap the Probe Setup panel.
If the Probe Compensation Status says Pass, the probe is already compensated for this channel. You can move the probe
to another channel and start again from step 1, or connect a different probe to this channel and start from step 1.
If the Probe Compensation Status says Default, continue with this procedure.
8. Tap Compensate Probe to open the Probe Compensation dialog.
9. Tap Compensate Probe to run the probe compensation.
10. The probe compensation is finished when the Probe Compensation Status displays Pass. Disconnect the probe tip and
ground from the PROBE COMP terminals.
11. Repeat these steps for each supported passive probe that you want to compensate for this channel.
12. Repeat these steps to compensate supported probes on other channels of the oscilloscope.
NOTE. For most accurate measurements, open the Probe Setup panel and verify the Probe Compensation Status is Pass
whenever you attach a probe to a channel.
Connect to a network (LAN)
Connecting to a network allows you to remotely access the instrument.
To access the user interface on the MSO58LP, connect a monitor to a video port on the rear of the instrument, and connect a
mouse to any USB Host port. You do not need to connect a mouse if your remote monitor is touch-capable. You can also
remotely access the user interface of a network-connected instrument by entering the instrument's IP address in a web browser.
To remotely set this control or run this task on an MSO58LP, see the 5 Series MSO MSO54, MSO56, MSO58, MSO58LP
Programmer Manual (Tektronix part number 077-1305-xx) for the correct command or commands to use.
Work with your network administrator to obtain the required information to connect to your network (IP address, Gateway IP
address, Subnet Mask, DNS IP address, and so on).
1. Connect a CAT5 cable from the oscilloscope LAN connector to your network.
2. Select Utility > I/O on the menu bar to open the I/O configuration menu.
Configure the instrument
14
MSO58LP Installation and Safety Manual
Summary of Contents for MSO 5 Series
Page 2: ......
Page 6: ......
Page 20: ...Installing your instrument 6 MSO58LP Installation and Safety Manual...
Page 24: ...Getting acquainted with your instrument 10 MSO58LP Installation and Safety Manual...
Page 30: ...Configure the instrument 16 MSO58LP Installation and Safety Manual...
Page 32: ...Operating basics 18 MSO58LP Installation and Safety Manual...
Page 34: ...Cleaning the instrument 20 MSO58LP Installation and Safety Manual...
Page 41: ...5 MSO MSO58LP P071356800 071 3568 00...
Page 42: ......
Page 43: ...5 MSO MSO58LP www tek com 071 3568 00 V1 4...
Page 45: ...3 a b c d W4 15AUG04...
Page 46: ......
Page 48: ...19 EMC 21 EMC 21 22 23 ii MSO58LP...
Page 49: ...MSO58LP iii...
Page 50: ...CAT 1 1 AC II III IV iv MSO58LP...
Page 51: ...CAT 2 2 BNC 0V MSO58LP v...
Page 52: ...vi MSO58LP...
Page 53: ...MSO58LP vii...
Page 54: ...viii MSO58LP...
Page 59: ...2 1 2 1 2 2 3 4 4 5 2 6 2 2 7 1 1Nm 8 5 16 2 4 0 7Nm 9 MSO58LP 3...
Page 60: ...AC AC 1 1 MSO58LP 2 AC AC Standby 3 AC 4 4 MSO58LP...
Page 62: ...TekVPI Menu TLP058 FlexChannel 1 2 FlexChannel 3 4 TLP058 BNC BNC 6 MSO58LP...
Page 74: ...18 MSO58LP...
Page 75: ...MSO58LP 19...
Page 76: ...20 MSO58LP...
Page 82: ...11 12 5 5 TPP0500 TPP1000 13 7 4 e Scope 16 Web 16 4 26 MSO58LP...
Page 83: ...5 MSO MSO58LP P071356800 071 3568 00...
Page 84: ......
Page 85: ...5 MSO MSO58LP www tek com 071 3568 00 V1 4...
Page 87: ...3 a b c d W4 15AUG04...
Page 88: ......
Page 90: ...19 EMC 21 EMC 21 22 23 ii MSO58LP...
Page 91: ...Tektronix CAT 1 1 II III IV MSO58LP iii...
Page 92: ...CAT iv MSO58LP...
Page 93: ...BNC 0 V MSO58LP v...
Page 94: ...DANGER WARNING CAUTION vi MSO58LP...
Page 99: ...1 2 3 4 5 6 7 1 1 N m 8 5 16 0 7 N m 9 MSO58LP 3...
Page 100: ...1 1 MSO58LP 2 3 4 4 MSO58LP...
Page 102: ...TLP058 FlexChannel 1 2 FlexChannel 3 4 TLP058 BNC BNC 6 MSO58LP...
Page 104: ...3 1kHz TPP0500B TPP1000 13 DUT ESD 4 USB USB U 5 Aux Trig SMA AUX In 6 LED 7 8 MSO58LP...
Page 106: ...10 MSO58LP...
Page 112: ...16 MSO58LP...
Page 114: ...18 MSO58LP...
Page 115: ...MSO58LP 19...
Page 116: ...20 MSO58LP...