Compensating a TPP0500B or TPP1000 passive voltage probe
Your Tektronix oscilloscope can automatically compensate TPP0500B and
TPP1000 probes. This eliminates the need for manual probe compensation, as is
typically performed with other probes.
Each compensation generates values for a specific probe and channel
combination. If you want to use the probe on another channel and desire to
compensate the new probe-channel pair, you must run a new set of compensation
steps for that new combination.
1.
Connect the oscilloscope power cable. See
on page 11.
2.
Power on the oscilloscope.
3.
Connect the probe connector to the oscilloscope channel, and the probe tip
and reference lead to the
PROBE COMP
terminals on the oscilloscope front
panel.
NOTE.
Connect only one probe at a time to the probe comp terminals.
4.
Push a front panel button for an input channel connected to the probe you
wish to compensate. (
1
,
2
,
3
, or
4
)
5.
Notice on the lower menu that the oscilloscope has automatically set the
probe termination value
Coupling
DC
| AC
Termination set
by TPP1000
Invert
On|
Off
Bandwidth
Full
Label
More
5
6
Installation
MDO4000C Series Oscilloscope User Manual
15
Preliminary