Trigger Setup
Starting and Stopping an Acquisition
After you have de
fi
ned the acquisition and trigger parameters, start the acquisition with
Run/Stop
or
Single
.
Push
Run/Stop
to start acquisitions.
The oscilloscope acquires repeatedly
until you push the button again to stop
the acquisition.
Push
Single
to take a single acquisition.
Single sets the trigger mode to
Normal
for the single acquisition.
If any analog or digital waveforms are
active, then pressing
Run/Stop
to stop
acquisitions will cause the oscilloscope
to wait for one more trigger event before
stopping. Once the acquisition occurs,
the
Run/Stop
button will turn red and the
Single
button will become unlit.
If the Trigger Mode is set to Auto and
another trigger event does not occur
within the Auto trigger timeout period
then an acquisition will be made and the
instrument will stop.
If the Trigger mode is set to Normal,
the oscilloscope will continue waiting
for a trigger event to occur as long as
necessary.
102
MDO3000 Series Oscilloscopes User Manual